Dose reduction with iterative reconstruction: Optimization of CT protocols in clinical practice - 07/05/15

Doi : 10.1016/j.diii.2015.02.007 
J. Greffier a, , F. Macri a, A. Larbi a, A. Fernandez a, E. Khasanova a, b, F. Pereira a, C. Mekkaoui a, c, J.P. Beregi a
a Department of Radiology, University Hospital Center of Nîmes, EA 2415, Bd Prof Robert-Debré, 30029 Nîmes cedex, France 
b Dana-Farber Cancer Institute, Harvard Medical School, Boston, USA 
c Harvard Medical School, Massachusetts General Hospital, Department of Radiology, Martinos Center for Biomedical Imaging, Boston, USA 

Corresponding author. CHU de Nîmes, Bd Prof Robert-Debré, 30029 Nîmes Cedex 9, France.

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Abstract

Objectives

To create an adaptable and global approach for optimizing MDCT protocols by evaluating the influence of acquisition parameters and Iterative Reconstruction (IR) on dose reduction and image quality.

Materials and methods

MDCT acquisitions were performed on quality image phantom by varying kVp, mAs, and pitch for the same collimation. The raw data were reconstructed by FBP and Sinogram Affirmed Iterative Reconstruction (SAFIRE) with different reconstruction kernel and thickness. A total of 4032 combinations of parameters were obtained. Indices of quality image (image noise, NCT, CNR, SNR, NPS and MTF) were analyzed. We developed a software in order to facilitate the optimization between dose reduction and image quality. Its outcomes were verified on an adult anthropomorphic phantom.

Results

Dose reduction resulted in the increase of image noise and the decrease of SNR and CNR. The use of IR improved these indices for the same dose without affecting NCT and MTF. The image validation was performed by the anthropomorphic phantom. The software proposed combinations of parameters to reduce doses while keeping indices of the image quality adequate. We observed a CTDIvol reduction between −44% and −83% as compared to the French diagnostic reference levels (DRL) for different anatomical localization.

Conclusion

The software developed in this study may help radiologists in selecting adequate combinations of parameters that allows to obtain an appropriate image with dose reduction.

El texto completo de este artículo está disponible en PDF.

Keywords : Multidetector CT, Iterative reconstruction, Dose reduction, Image quality, Patient safety

Abbreviations : CNR, CTDI, DRL, FBP, FOV, IR, LDPE, LSF, MDCT, MTF, NCT, NPS, PSF, ROI, SAFIRE, SNR, VBA


Esquema


© 2015  Éditions françaises de radiologie. Publicado por Elsevier Masson SAS. Todos los derechos reservados.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 96 - N° 5

P. 477-486 - mai 2015 Regresar al número
Artículo precedente Artículo precedente
  • Analysis of subsequent publication of scientific orally presented abstracts of the French National Congress of Radiology. Part II: Focus on the French abstracts
  • V. Dangouloff-Ros, M. Ronot, M. Lagadec, V. Vilgrain
| Artículo siguiente Artículo siguiente
  • CT pulmonary angiogram with 60% dose reduction: Influence of iterative reconstructions on image quality
  • F. Pontana, D. Moureau, B. Schmidt, A. Duhamel, J.-B. Faivre, K. Yasunaga, J. Remy, M. Remy-Jardin

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

@@150455@@ Voir plus

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2026 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.