Précision réfractive et qualité de vision objective après implantation torique pseudophaque - 09/10/12

Doi : 10.1016/j.jfo.2011.10.012 
A. Debois a, , Y. Nochez a, b, C. Bezo a, D. Bellicaud a, b, P.-J. Pisella a, b
a Service d’ophtalmologie, CHU de Bretonneau, 2, boulevard Tonnellé, 37000 Tours, France 
b Faculté de médecine François-Rabelais, 59, rue Néricault-Destouches, 37000 Tours, France 

Auteur correspondant.

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
Articolo gratuito.

Si connetta per beneficiarne

Résumé

Introduction

Étudier l’efficacité d’une implantation torique pseudophaque pour corriger un astigmatisme cornéen préopératoire en analysant la précision réfractive sphérocylindrique et la qualité de vision objective postopératoire.

Patients et méthodes

Étude prospective de 13 yeux bénéficiant d’une chirurgie micro-incisionnelle de la cataracte à 1,8mm avec implantation torique (Lentis L313T®, Oculentis) afin de compenser un astigmatisme cornéen préopératoire supérieur à un dioptrie. L’évaluation préopératoire comprend une kératométrie, une réfraction subjective et une aberrométrie totale et cornéenne (KR-1®, Topcon). Les mesures postopératoires à six mois comprennent une photographie de l’implant (rotation et décentrement), une mesure de l’acuité visuelle sans et avec correction, une mesure objective de la qualité de vision (OQAS®, Visiometrics, Espagne).

Résultats

En postopératoire, l’acuité visuelle sans correction est de 8,33/10e±1,91 (0,09±0,11 LogMar). La sphère réfractive moyenne postopératoire est de 0,13±0,73 dioptries. L’astigmatisme final est de –0,66±0,56 dioptries pour un astigmatisme cornéen de 2,17±0,68 dioptries. La rotation moyenne d’axe de l’implant torique est de 4,4°±3,69° (entre 0°et 10°).

Discussion

La rotation moyenne de cet implant à six mois est inférieure à 5° démontrant la stabilité de l’optique dans le sac capsulaire. Par ailleurs, les mesures objectives de qualité de vision confirment les résultats subjectifs d’acuité visuelle sans correction.

Conclusion

L’implantation pseudophaque par l’implant asphérique torique L313T® est efficace pour la correction d’un astigmatisme cornéen lors d’une chirurgie micro-incisionnelle de 1,8mm.

Il testo completo di questo articolo è disponibile in PDF.

Summary

Purpose

To study efficacy and predictability of toric IOL implantation for correction of preoperative corneal astigmatism by analysing spherocylindrical refractive precision and objective quality of vision.

Patients and methods

Prospective study of 13 eyes undergoing micro-incisional cataract surgery through a 1.8mm corneal incision with toric IOL implantation (Lentis L313T®, Oculentis) to treat over one D of preoperative corneal astigmatism. Preoperative evaluation included keratometry, subjective refraction, and total and corneal aberrometry (KR-1®, Topcon). Six months postoperatively, measurements included slit lamp photography, documenting IOL rotation, tilt or decentration, uncorrected visual acuity, best-corrected visual acuity and objective quality of vision measurement (OQAS® Visiometrics, Spain).

Results

Postoperatively, mean uncorrected distance visual acuity was 8.33/10±1.91 (0.09±0.11 LogMar). Mean postoperative refractive sphere was 0.13±0.73 diopters. Mean refractive astigmatism was –0.66±0.56 diopters with corneal astigmatism of 2.17±0.68 diopters. Mean IOL rotation was 4.4°±3.6° (range 0° to 10°).

Discussion

Mean rotation of this IOL at 6 months was less than 5°, demonstrating stability of the optic within the capsular bag. Objective quality of vision measurements were consistent with subjective uncorrected visual acuity.

Conclusion

Implantation of the L313T® IOL is safe and effective for correction of corneal astigmatism in 1.8mm micro-incisional cataract surgery.

Il testo completo di questo articolo è disponibile in PDF.

Mots clés : Astigmatisme, Chirurgie de la cataracte, Précision réfractive, Qualité de vision

Keywords : Astigmatism, Cataract surgery, Refractive precision, Quality of vision


Mappa


© 2012  Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 35 - N° 8

P. 580-586 - Ottobre 2012 Ritorno al numero
Articolo precedente Articolo precedente
  • Un nouveau cas de trou maculaire géant chez un patient présentant un syndrome d’Alport
  • E. Mercé, J.-F. Korobelnik, M.-N. Delyfer, M.-B. Rougier
| Articolo seguente Articolo seguente
  • Relations entre les paramètres de la réfraction : sphère, cylindre et axe
  • O. Touzeau, T. Gaujoux, J. Bullet, C. Allouch, V. Borderie, L. Laroche

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2024 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.