Radiation Exposure to the Urologist Using an Overcouch Radiation Source Compared With an Undercouch Radiation Source in Contemporary Urology Practice - 11/04/18

Doi : 10.1016/j.urology.2017.12.011 
Andrew M. Harris *
 Department of Urology, University of Kentucky Medical Center, Lexington, KY 

*Address correspondence to: Andrew M. Harris, M.D., University of Kentucky Medical Center, 800 Rose St., MS 237, Lexington, KY.University of Kentucky Medical Center800 Rose St., MS 237LexingtonKY

Abstract

Objective

To compare radiation dosage to the urologist using an overcouch system, x-ray tube over table, and an undercouch system, x-ray tube under table. Urologists continue to perform more endoscopic surgery requiring fluoroscopy. Fluoroscopy, or electromagnetic radiation, can cause cellular damage when passing through tissues. These systems are compared with respect to radiation dosage to the urologist.

Methods

A single urologic surgeon utilized a dosimeter badge while using an overcouch system. The dosimeter exposure was higher than expected and an undercouch system was then employed. Dosimeter exposure levels between the overcouch and the undercouch systems were examined and compared.

Results

Over the 4 months reviewed for the overcouch system, radiation doses to the body averaged 3.63 mSv, those to the eye averaged 3.73 mSv, and those to the extremities averaged 3.72 mSv. The 3-month averages for the undercouch system exposure to the body, the eye, and the extremities were 0.31, 0.35, and 0.35 mSv, respectively. The difference in radiation exposure between the 2 systems was significant (P ≤.001). The average number of radiation cases between the 2 systems was not significantly different (P = .37). The average fluoroscopy time for the procedures between the 2 systems was not significantly different (P = .24).

Conclusion

Overcouch fluoroscopy systems expose the urologist to significantly higher, potentially dangerous levels of radiation. Urologists using an overcouch system should strongly consider as low as reasonably achievable precautions and proper utilization of lead aprons, thyroid shields, and lead glasses. Radiation safety training should be considered.

Le texte complet de cet article est disponible en PDF.

Plan


 Financial Disclosure: The authors declare that they have no relevant financial interests.


© 2017  Elsevier Inc. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 114

P. 45-48 - avril 2018 Retour au numéro
Article précédent Article précédent
  • Emergency Department Switching and Duplicate Computed Tomography Scans in Patients With Kidney Stones
  • Parth K. Shah, Phyllis L. Yan, Casey A. Dauw, Brent K. Hollenbeck, Khurshid R. Ghani, Amy N. Luckenbaugh, John M. Hollingsworth
| Article suivant Article suivant
  • Effect of Intercostal Nerve Block and Nephrostomy Tract Infiltration With Ropivacaine on Postoperative Pain Control After Tubeless Percutaneous Nephrolithotomy: A Prospective, Randomized, and Case-controlled Trial
  • Sae Woong Choi, Shin Jay Cho, Hyong Woo Moon, Kyu Won Lee, Sang Hoon Lee, Sang Hyun Hong, Yong Sun Choi, Woong Jin Bae, U-Syn Ha, Sung-Hoo Hong, Ji Youl Lee, Sae Woong Kim, Hyuk Jin Cho

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.