Échographie destendons de la coiffe en mode harmonique : les clivages sont-ilsenfin visibles ?

Objectifs. Pourcertains, les clivages restent un des points faibles de l’échographie.Le but de cette étude est de démontrer la visibilité et décrirela séméiologie des clivages des tendons supraet infra-épineux en mode harmonique.

Matérielset methods. Il s’agit d’une étudeprospective de 52 patients suspects de lésionstendineuses ayant bénéficié d’une échographieavant un arthroscanner, un bursoscanner ou une arthro-IRM. L’examenest réalisé avec une sonde 7-15 MHZ enmode harmonique tissulaire (« pulse substraction »).Un clivage était évoqué devant la miseen évidence d’une ligne hypoéchogèneintratendineuse, prolongeant une rupture partielle ou transfixiante.Les performances de l’échographie concernant lesclivages étaient corrélées aux examensopacifiés.

Résultats. Dix cas declivages étaient décrits en échographiecontre 18 visibles sur les examens arthrographiques. Les faux négatifs étaienten rapport avec des patients peu échogènes (n = 4),des tendons trop rétractés (n = 2)ou opérés (n = 2). La sensibilité étaitde 55 % et la spécificité de94 %, la valeur prédictive positive de83 %, la valeur prédictive négativede 80 % pour le dépistage des clivages.Les performances de l’échographie étaientmoins bonnes pour les clivages associés à des rupturestransfixiantes (5/11) en comparaison aux clivages associésaux ruptures partielles (5/6).

Conclusion. Nousdémontrons que les clivages sont maintenant visibles en échographiegrâce aux sondes hautes fréquences et au modeharmonique tissulaire. Toutefois, la sensibilité resteinsuffisante surtout en cas de lésion transfixiante évoluée,en postopératoire et pour les patients peu échogènes.

Purpose. For some, cleavage tearsremain a pitfall of sonography (US). The purpose of this study isto demonstrate the visibility of intratendinous tears of the supraspinatusand infraspinatus tendons and describe their imaging features ontissue harmonic US.

Materials and methods. Prospectivestudy of 52 patients with suspected cuff pathology who underwentUS prior to CT-arthrography, CT-bursography or MR arthrography.The US examinations were performed using 7-15MHz transducers withtissue harmonic mode (pulse subtraction). An intratendinous tearwas suggested by the presence of a hypoechoic intratendinous line, extendingfrom a partial or full thickness tear. Results from US were correlatedto contrast material enhanced CT or MR findings.

Results. Tencases of cleavage tears were detected on US compared to 18 on arthrographicexaminations. False negative results occurred in poor US candidates(n=4), excessively retracted tendons (n=2) or postsurgicalcuffs (n=2). Sensitivity was 55% and specificitywas 94%, with PPV of 83% and NPV of 80% forthe detection of intratendinous tear. The accuracy of US was lowerfor intratendinous tear associated with full thickness tears (5/11)compared to intratendinous tears associated with partial thicknesstears (5/6).

Conclusion. We demonstrate thatcleavage tears are now visible on US using high-frequency transducersand tissue harmonic mode. However, the sensitivity remains too low,especially in patients with full thickness tear, postsurgical patientsand patients that are poor candidates to US.


Mots clés : Épaule , Échographie , Tendons , Rupture , Arthrographie , IRM , Scanner

Keywords: Shoulder , US , Tendons , Tear , Arthrography , MRI , CT


Plan



© 2008 Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 89 - N° 3-C1

P. 333-338 - mars 2008 Retour au numéro

Bienvenue sur EM-consulte, la référence des professionnels de santé.

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.