Recommandations de la Société d’imagerie abdominale et digestive pour l’utilisation des produits de contraste iodés en tomodensitométrie abdominale et digestive : protocoles selon l’indication et optimisation des volumes - 25/09/26

French Society of Abdominal and Digestive Imaging recommendations for the use of iodinated contrast media in abdominal and digestive computed tomography: Protocols according to indication and volume optimization

Doi : 10.1016/j.jidi.2026.04.005 
J. Grégory a, ⁎ , G. Simon b, V. Laurent c, M. Lewin d, M. Wagner e, M. Zappa f, M. Ronot a, P. Rousset g
a Service de radiologie, hôpital Beaujon, AP–HP Nord, FHU MOSAIC, centre de recherche sur l’inflammation (CRI), Inserm U1149, université Paris Cité, 100, boulevard du Général-Leclerc, 92110 Clichy, France 
b Service de radiologie, CHRU de Besançon, université de Bourgogne Franche-Comté, 3, boulevard Alexandre-Fleming, 25030 Besançon cedex, France 
c Service de radiologie adulte, IADI, U12454, Inserm, CHRU de Nancy, université de Lorraine, rue du Morvan, 54511 Vandœuvre-les-Nancy, France 
d Service de radiologie, hôpital Paul-Brousse, AP–HP, université Paris-Saclay, 12, avenue Paul-Vaillant-Couturier, 94800 Villejuif, France 
e Service de radiologie, hôpital Saint-Antoine, Sorbonne université, AP–HP, 184, rue du Faubourg-Saint-Antoine, 75012 Paris, France 
f Service de radiologie, CH de Cayenne Andrée-Rosemon, rue des flamboyants, 97306 Cayenne, Guyane française 
g Service d’imagerie diagnostique et interventionnelle, CICLY-EA3738, hôpital Lyon Sud, hospices civils de Lyon, université Claude-Bernard Lyon 1, 165, chemin du Grand-Revoyet, 69495 Pierre-Bénite, France 

⁎ Auteur correspondant : service de radiologie, hôpital Beaujon, AP–HP Nord, FHU MOSAIC, centre de recherche sur l’inflammation (CRI), Inserm U1149, université Paris Cité, 100, boulevard du Général-Leclerc, 92110 Clichy, France. service de radiologie, hôpital Beaujon, AP–HP Nord, FHU MOSAIC, centre de recherche sur l’inflammation (CRI), Inserm U1149, université Paris Cité 100, boulevard du Général-Leclerc Clichy 92110 France

Résumé

Introduction

L’utilisation des produits de contraste iodés (PCI) en tomodensitométrie (TDM) abdominale et digestive est quotidienne et répond à des objectifs d’opacification vasculaire et d’analyse tissulaire. En l’absence de recommandations consolidées, la Société d’imagerie abdominale et digestive (SIAD), sous l’égide de la Société française de radiologie (SFR), a élaboré des directives visant à standardiser les protocoles d’acquisition et à optimiser le volume de PCI administré.

Messages principaux

Des protocoles d’acquisition sont définis pour chaque grande catégorie d’indication, précisant les phases (sans injection, artérielle vasculaire, artérielle tardive, pancréatique, portale, tardive) à réaliser en fonction du contexte clinique. L’optimisation du volume injecté repose sur plusieurs leviers complémentaires : modulation en fonction du poids corporel de la charge en Iode (charge minimale de 0,5 g d’iode/kg), réduction du kilovoltage pour augmenter la visibilité de l’iode, usage de l’imagerie spectrale avec reconstruction à basse énergie, optimisation du débit d’injection et utilisation d’un bolus de rinçage. L’usage multipatient, désormais possible dans un cadre réglementaire défini (arrêté du 1 er juin 2023), et les circuits de recyclage contribuent à réduire l’empreinte environnementale de l’activité d’imagerie.

Conclusion

La réduction du volume de PCI en TDM abdominale est réalisable sans sacrifice de la qualité diagnostique, à condition de respecter des seuils de rehaussement définis et d’utiliser les techniques d’optimisation disponibles. La résolution en contraste reste la priorité absolue.

Le texte complet de cet article est disponible en PDF.

Summary

Introduction

The use of iodinated contrast media (ICM) in abdominal and digestive computed tomography (CT) is routine and serves two objectives: vascular opacification and tissue contrast. The Société d’imagerie abdominale et digestive (SIAD), under the auspices of the Société française de radiologie (SFR), developed recommendations to standardize acquisition protocols and optimize ICM volumes in the absence of consolidated guidelines.

Main messages

Acquisition protocols are defined for each major indication category, specifying which phases should be performed (unenhanced, early arterial, late arterial, pancreatic, portal venous, delayed). ICM volume optimization relies on several complementary strategies: weight-based dose calculation (minimum iodine load 0.5 gI/kg), kilovoltage reduction to enhance iodine visibility, dual-energy CT with low-energy reconstruction (40 keV), injection flow rate optimization, and saline flush. Multivial use, now permitted under French regulation (decree of June 1, 2023), and recycling circuits further reduce the environmental footprint.

Conclusion

ICM volume reduction in abdominal CT is achievable without compromising diagnostic quality, provided that defined enhancement thresholds are met and optimization techniques are applied. Contrast resolution remains the absolute priority.

Le texte complet de cet article est disponible en PDF.

Mots clés : Tomodensitométrie abdominale, Produits de contraste iodés, Protocoles d’acquisition, Optimisation des doses, Imagerie spectrale

Keywords : Abdominal CT, Iodinated contrast media, Acquisition protocols, Dose optimization, Spectral imaging


Plan


© 2026  The Authors. Publié par Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 9 - N° 5

P. 227-233 - octobre 2026 Retour au numéro
Article précédent Article précédent
  • Masses rénales indéterminées : quand l’échographie de contraste fait la différence
  • S. Bodard, G. Reichert, C. Tescher, J.-M. Correas, O. Hélénon, A. O’Keane
| Article suivant Article suivant
  • Recommandations de la Société d’imagerie abdominale et digestive pour l’utilisation des produits de contraste gadoliniés en IRM abdominale et digestive : indications, choix de l’agent et pratiques durables
  • J. Grégory, G. Simon, M. Wagner, M. Lewin, V. Laurent, M. Zappa, M. Ronot, P. Rousset

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Elsevier s'engage à rendre ses eBooks accessibles et à se conformer aux lois applicables. Compte tenu de notre vaste bibliothèque de titres, il existe des cas où rendre un livre électronique entièrement accessible présente des défis uniques et l'inclusion de fonctionnalités complètes pourrait transformer sa nature au point de ne plus servir son objectif principal ou d'entraîner un fardeau disproportionné pour l'éditeur. Par conséquent, l'accessibilité de cet eBook peut être limitée. Voir plus

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2026 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.