Optimising accuracy of computer-assisted implant surgery using novel drilling key design: simulation-based study - 16/10/26

Doi : 10.1016/j.jormas.2026.103020 
Thaysan Kaing a, b, Sirida Arunjaroensuk a, b, Xin Hui Yeo a, b, Min Thu Kha Ko Ko a, b, Nikos Mattheos a, b, c, Atiphan Pimkhaokham a, b, ⁎
a Oral and Maxillofacial surgery and Digital Implant Surgery Research Unit, Faculty of Dentistry, Chulalongkorn University, Bangkok, Thailand 
b Department of Oral and Maxillofacial Surgery, Faculty of Dentistry, Chulalongkorn University, Bangkok, Thailand 
c Department of Dental Medicine, Karolinska Institute, Stockholm, Sweden 

⁎ Corresponding author at: Department of Oral and Maxillofacial Surgery, Faculty of Dentistry, Chulalongkorn University, 34 Henri Dunant Road, Wangmai, Patumwan, Bangkok, 10330, Thailand. Department of Oral and Maxillofacial Surgery, Faculty of Dentistry Chulalongkorn University 34 Henri Dunant Road, Wangmai, Patumwan Bangkok 10330 Thailand

Abstract

Background

Accurate implant positioning remains the goal of guided surgery for improving surgical outcomes and minimizing complications. The aim is to compare the accuracy of standard versus novel drilling key in static-dynamic computer assisted implant surgery across D2 and D3 bone densities relative to a zero-deviation benchmark.

Methods

Sixty-four bone-level tapered implants were placed in 32 models across four groups (n = 16), stratified by drilling key designs (standard vs novel) and bone density (D2 vs D3). Preoperative cone-beam computed tomography (CBCT) was imported into the navigation system for implant planning. Surgical guide was designed to match the navigation planning. Implant placement followed the manufacturer protocols. Postoperative CBCT was superimposed on pre-operative plan to measure 3D platform, 3D apex, and angular deviations. Two-way ANOVA, Levene, Kruskal–Wallis, and post-hoc Mann–Whitney U tests were used ( p < 0.05).

Results

The novel drilling key significantly reduced angular deviation compared to the standard key (1.84°± 1.58 vs 0.21°± 0.56, p < 0.001); 3D platform (0.78 ± 0.30 vs 0.74 ± 0.17 mm, p = 0.262) and apex (0.82 ± 0.31 vs 0.93 ± 0.41 mm , p = 0.596) were comparable. D3 bone showed higher 3D apex ( p = 0.017) and angular deviations ( p = 0.044) than D2.

Conclusion

The findings imply that the novel key design provides a targeted advantage in reducing angular deviation in both bone densities while linear deviations at the platform and apex remain influenced by bone density and other potential sources of error within the guided workflow.

Clinical significance

This study addresses accuracy in implant dentistry in technically challenging cases such as long free drilling distance where low sleeve positioning is not feasible.

Le texte complet de cet article est disponible en PDF.

Keywords : sd-CAIS, Sleeve-in-sleeve surgical guide, Novel drilling key design, Navigation system, Bone density, And implant accuracy


Plan


© 2026  Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 127 - N° 6

Article 103020- décembre 2026 Retour au numéro
Article précédent Article précédent
  • Clinical outcomes of guided bone regeneration in anterior maxilla with labial dehiscence: a retrospective cohort study
  • Danji Zhu, Mengdie Fu, Wenyi Xu, Qinmeng Zhang, Yan Chen, Jing Zhang, Zhiwei Jiang, Guoli Yang
| Article suivant Article suivant
  • Association of night-time stabilization splint use with condylar and posterior eminence remodeling in adolescents with unilateral temporomandibular joint osteoarthritis: a CBCT-based retrospective study
  • Jae Joon Hwang, Hye-Mi Jeon, Hye-Min Ju, Sung-Hee Jeong, Yong-Woo Ahn, Soo-Min Ok

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Elsevier s'engage à rendre ses eBooks accessibles et à se conformer aux lois applicables. Compte tenu de notre vaste bibliothèque de titres, il existe des cas où rendre un livre électronique entièrement accessible présente des défis uniques et l'inclusion de fonctionnalités complètes pourrait transformer sa nature au point de ne plus servir son objectif principal ou d'entraîner un fardeau disproportionné pour l'éditeur. Par conséquent, l'accessibilité de cet eBook peut être limitée. Voir plus

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2026 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.