Doi : 10.1016/j.otsr.2026.104983 
Charles Rivière a, b, ⁎ , Giuseppe Aloisi a, c, Loïc Villet a, b
a Bordeaux Arthroplasty Research Institute (BARI), 6 rue Georges Negrevergne, 33700 Mérignac, France 
b Clinique du Sport Bordeaux Mérignac, 4 rue Georges Negrevergne, 33700 Mérignac, France 
c Università Degli Studi Dell’Aquila, Department of Life, Health and Environmental Sciences, 67100 L’Aquila, Italy 

⁎ Corresponding author.
Sous presse. Épreuves corrigées par l'auteur. Disponible en ligne depuis le Saturday 07 November 2026

Abstract

Purpose

Accurate tibial resection is critical in unrestricted kinematic alignment total knee arthroplasty (uKA-TKA), as component positioning aims to restore the native joint line through anatomical resurfacing. This study compared the reproducibility of initial tibial resections performed with manual instrumentation versus patient-specific instrumentation (PSI) in uKA-TKA. The secondary objective was to compare tibial recut rates between techniques.

Methods

This retrospective comparative cohort study analyzed prospectively collected intraoperative measurements from 124 primary uKA-TKAs performed between June 2024 and January 2026. Tibial resections were performed using manual instrumentation in 83 cases and PSI in 41 cases. Target and measured tibial resection thicknesses were recorded intraoperatively at the medial (MTP) and lateral tibial plateaus (LTP) using an analog sliding caliper (0.5-mm resolution). Resection error (Δ) was defined as measured minus target thickness. Accuracy was expressed as mean signed error and precision as the standard deviation of signed error. Absolute error thresholds and tibial recut rates were also analyzed.

Results

Mean signed error did not differ significantly between techniques for either plateau (medial: 0.15 mm PSI vs −0.10 mm manual, p = 0.158; lateral: 0.10 mm PSI vs −0.05 mm manual, p = 0.378), indicating comparable average accuracy. However, medial precision was significantly greater with manual instrumentation (SD 0.64 mm vs 1.02 mm with PSI, p = 0.002), with lower mean absolute error (0.38 mm vs 0.73 mm; p = 0.007). Fewer medial resections exceeded the ±1.5 mm threshold in the manual group (2.4% vs 9.8%, p = 0.024). No significant differences were observed laterally. Angular tibial recut was performed in 12.0% of manual cases and 22.0% of PSI cases (p = 0.187).

Conclusion

Manual instrumentation and PSI achieved similar mean accuracy for initial tibial resections in uKA-TKA. However, manual instrumentation was associated, in this two-surgeon series, with greater medial reproducibility and fewer large resection deviations, with a numerically lower, non-significant recut rate. Within the sequential uKA workflow – including caliper verification and corrective recut when required – both techniques allow reliable convergence toward the predefined resurfacing target.

Level of evidence

III; retrospective comparative study.

Le texte complet de cet article est disponible en PDF.

Keywords : Kinematic alignment, Total knee arthroplasty, Bone resection accuracy, Patient-specific instrumentation, Caliper-verified technique

Abbreviations : KA, TKA, uKA TKA, MTP, LTP, PSI, CT, SD, CPAK


Plan


© 2026  Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Elsevier s'engage à rendre ses eBooks accessibles et à se conformer aux lois applicables. Compte tenu de notre vaste bibliothèque de titres, il existe des cas où rendre un livre électronique entièrement accessible présente des défis uniques et l'inclusion de fonctionnalités complètes pourrait transformer sa nature au point de ne plus servir son objectif principal ou d'entraîner un fardeau disproportionné pour l'éditeur. Par conséquent, l'accessibilité de cet eBook peut être limitée. Voir plus

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2026 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.