Correspondence between tidal volume and tidal impedance variation in intubated patients - 20/11/26

Doi : 10.1016/j.aicoj.2026.100210 
Vincent Joussellin a, b, i, Pascale Labedade c, d, e, François Arrive f, Mathilde Taillantou-Candau g, François Beloncle g, Rémi Coudroy f, h, Guillaume Carteaux c, d, e, Martin Dres i, j, ⁎
a CHU Rennes, Maladies Infectieuses et Réanimation Médicale, F-35033 Rennes, France 
b Université Rennes, Faculté de Médecine, Biosit, F-35043 Rennes, France 
c Assistance Publique-Hôpitaux de Paris, CHU Henri Mondor, Créteil, France 
d Université Paris Est-Créteil, Groupe de Recherche Clinique CARMAS, Créteil, France 
e Institut Mondor de Recherche Biomédicale INSERM 955, Créteil, France 
f Médecine Intensive Réanimation, Centre Hospitalier Universitaire de Poitiers, 2 rue la Milétrie, 86021, Poitiers Cedex, France 
g Medical Intensive Care Unit (ICU), Angers University Hospital, Vent'Lab, University of Angers, Angers, France 
h INSERM CIC 1402, IS-ALIVE Research group, University of Poitiers, Poitiers, France 
i Sorbonne Université, INSERM, UMRS1158 Neurophysiologie respiratoire expérimentale et clinique, Paris, France 
j AP-HP. Sorbonne Université, Hôpital Pitié-Salpêtrière, Service de Médecine Intensive – Réanimation (Département "R3S"), F-75013, Paris, France 

⁎ Corresponding author.

Bienvenue sur EM-consulte, la référence des professionnels de santé.
Article gratuit.

Connectez-vous pour en bénéficier!

Sous presse. Manuscrit accepté. Disponible en ligne depuis le Friday 20 November 2026

Highlights

•
The ratio between tidal volume and tidal impedance variation varies considerably between critically ill patients when assessed separately within each EIT device cohort (PulmoVista or Enlight).
•
Body mass index is strongly associated with TIV/VT ratio variation.
•
Accurate tidal volume estimation from EIT data without individual calibration does not appear feasible.
•
The large TIV/VT ratio intraindividual variability observed over time represents an inherent limitation of the calibration approach.

Le texte complet de cet article est disponible en PDF.

Abstract

Background

During mechanical ventilation, patient characteristics may affect the correlation between tidal volume (VT) and tidal impedance variation (TIV) measured by electrical impedance tomography (EIT). This study aimed at quantifying the TIV/VT ratio – the multiplicative coefficient used to estimate tidal volume from tidal impedance variation – across various settings of acute respiratory failure, and to explore its determinants.

Methods

Adults mechanically ventilated through an orotracheal tube who had a PaO₂/FiO₂ ratio below 300 mmHg were included across four centers in France. Tidal impedance variation was measured using two distinct EIT devices (PulmoVista 500® (Dräger, Germany) and Enlight 1800® (Timpel, Brazil)) while expired tidal volume was measured by the ventilator or a pneumotachograph. TIV/VT ratio was defined as the ratio between tidal volume and tidal impedance variation for each respiratory cycle during the recording period, then averaged.

Results

120 patients were included in the study: 74 (62 %) monitored with the PulmoVista device and 46 (38%) with the Enlight device. The TIV/VT ratio exhibited substantial interindividual variability, with median values of 0.17 (IQR 0.14 – 0.26) mL/AU for the PulmoVista device and 22.8 (IQR 17.7 – 26.7) mL/AU for the Enlight device, and intraindividual variability, with a median coefficient of variation of 10 (IQR 5 – 23) %. Using stepwise multiple linear regression, the final model for the TIV/VT ratio retained body mass index, plasma protein concentration and hematocrit for the PulmoVista device (adjusted R² = 0.36, p < 0.001), and body mass index, airway resistance, temperature, hematocrit, and heart rate for the Enlight device (adjusted R² = 0.63, p   <   0.001).

Conclusions

The impedance-to-tidal volume relationship displays substantial interindividual variability in critically ill patients, making individual calibration mandatory for tidal volume estimation. In addition, the large intraindividual variability observed over time represents an inherent limitation of this approach.

Le texte complet de cet article est disponible en PDF.

Keywords : Electrical impedance tomography, mechanical ventilation, tidal volume, tidal impedance variation

Abbreviations : AIC, ARDS, AU, BMI, CV, EELI, EIT, FiO2, ICU, IQR, PaCO 2 , PaO 2 , PBW, PEEP, SAPS II, SOFA, TIV, ΔZ, VT, VIF


Plan


© 2026  Publié par Elsevier Masson SAS.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Bienvenue sur EM-consulte, la référence des professionnels de santé.

Elsevier s'engage à rendre ses eBooks accessibles et à se conformer aux lois applicables. Compte tenu de notre vaste bibliothèque de titres, il existe des cas où rendre un livre électronique entièrement accessible présente des défis uniques et l'inclusion de fonctionnalités complètes pourrait transformer sa nature au point de ne plus servir son objectif principal ou d'entraîner un fardeau disproportionné pour l'éditeur. Par conséquent, l'accessibilité de cet eBook peut être limitée. Voir plus

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2026 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.