Correspondence between tidal volume and tidal impedance variation in intubated patients - 20/11/26

Highlights |
• | The ratio between tidal volume and tidal impedance variation varies considerably between critically ill patients when assessed separately within each EIT device cohort (PulmoVista or Enlight). |
• | Body mass index is strongly associated with TIV/VT ratio variation. |
• | Accurate tidal volume estimation from EIT data without individual calibration does not appear feasible. |
• | The large TIV/VT ratio intraindividual variability observed over time represents an inherent limitation of the calibration approach. |
Abstract |
Background |
During mechanical ventilation, patient characteristics may affect the correlation between tidal volume (VT) and tidal impedance variation (TIV) measured by electrical impedance tomography (EIT). This study aimed at quantifying the TIV/VT ratio – the multiplicative coefficient used to estimate tidal volume from tidal impedance variation – across various settings of acute respiratory failure, and to explore its determinants.
Methods |
Adults mechanically ventilated through an orotracheal tube who had a PaO₂/FiO₂ ratio below 300 mmHg were included across four centers in France. Tidal impedance variation was measured using two distinct EIT devices (PulmoVista 500® (Dräger, Germany) and Enlight 1800® (Timpel, Brazil)) while expired tidal volume was measured by the ventilator or a pneumotachograph. TIV/VT ratio was defined as the ratio between tidal volume and tidal impedance variation for each respiratory cycle during the recording period, then averaged.
Results |
120 patients were included in the study: 74 (62 %) monitored with the PulmoVista device and 46 (38%) with the Enlight device. The TIV/VT ratio exhibited substantial interindividual variability, with median values of 0.17 (IQR 0.14 – 0.26) mL/AU for the PulmoVista device and 22.8 (IQR 17.7 – 26.7) mL/AU for the Enlight device, and intraindividual variability, with a median coefficient of variation of 10 (IQR 5 – 23) %. Using stepwise multiple linear regression, the final model for the TIV/VT ratio retained body mass index, plasma protein concentration and hematocrit for the PulmoVista device (adjusted R² = 0.36, p < 0.001), and body mass index, airway resistance, temperature, hematocrit, and heart rate for the Enlight device (adjusted R² = 0.63, p < 0.001).
Conclusions |
The impedance-to-tidal volume relationship displays substantial interindividual variability in critically ill patients, making individual calibration mandatory for tidal volume estimation. In addition, the large intraindividual variability observed over time represents an inherent limitation of this approach.
Le texte complet de cet article est disponible en PDF.Keywords : Electrical impedance tomography, mechanical ventilation, tidal volume, tidal impedance variation
Abbreviations : AIC, ARDS, AU, BMI, CV, EELI, EIT, FiO2, ICU, IQR, PaCO 2 , PaO 2 , PBW, PEEP, SAPS II, SOFA, TIV, ΔZ, VT, VIF