Measurement Calibration of Two-Dimensional DSA Images When the Target Deviates from the Isocenter - 23/11/26
, Yuan Dong a, Lubin Yan a, Xiuli Zhao bAbstract |
Measurement accuracy in two-dimensional digital subtraction angiography (2D-DSA) images is compromised when the target deviates from the isocenter. This study aimed to derive simplified calibration equations for non-isocentric scenarios and to validate their accuracy.
Using two Siemens single-plane DSA systems (Artis Pheno and Artis Zee), a 10-mm steel sphere was measured at non-isocentric positions across multiple C-arm angulations and table movements. Sphere diameter was measured on the DSA workstation, and errors were compared before and after calibration. Using the Artis Zee, seven steel spheres were arranged in a plane parallel to the detector and offset from the isocenter. Images were acquired in frontal and lateral projections at different table positions and magnification fields of view (FOVs). Measured diameters were compared across positions and FOVs.
For the single sphere imaged at multiple angulations and table movements, pre-calibration maximum errors were 22.05% and 24.70% for the two systems; after calibration, maximum errors were reduced to 3.18% and 2.15%, respectively. The standard deviations of the paired measured and calibrated values differed significantly for both systems (both P < 0.001). For the seven-sphere array imaged in frontal and lateral projections, no statistically significant difference in measured diameter was observed across different positions ( χ 2 = 0.88, P = 0.990 ) or across different magnification FOVs ( χ 2 = 2.32, P = 0.803 ).
The proposed non-isocentric calibration method significantly improves measurement accuracy in 2D-DSA imaging, with important implications for neurointerventional procedures requiring precise vessel measurements.
Le texte complet de cet article est disponible en PDF.Graphical abstract |
Highlights |
• | Non-isocentric targets without calibration cause errors in 2D-DSA measurements. |
• | The calibration equations were derived using DICOM parameters. |
• | Maximum error decreased from >24% to <3.2% after calibration. |
• | The method is valid for various C-arm angulations and table positions. |
• | It improves the reliability of vessel sizing in neurointerventional procedures. |
Keywords : Digital subtraction angiography, Two-dimensional image, Isocenter, Non-isocentric, Calibration, Accurate measurement