S'abonner

Randomized comparison of mounted versus unmounted stents: The multicenter COMUS trial - 28/08/11

Doi : 10.1067/mhj.2003.117 
Thorsten I. Schneider, MDa, Hans W. Höpp, MDa, Dominik Vlaho, MSa, Gernot Wassmer, PhDb, Rolf Füssl, MDc, Sven Fassbender, MDd, Gabriele Wehr, MDe, Karl H. Konz, MDf, Frank Späh, MDg, Frank M. Baer, MDa
Cologne, Ulm, Frankfurt, Tübingen, Mönchengladbach, and Krefeld, Germany 
From the aDepartment III Internal Medicine, bIMSI Institute for Medical Statistics, Informatics and Epidemiology, University of Cologne, Cologne, cDepartment II Internal Medicine, University of Ulm, Ulm, dDepartment IV Internal Medicine, University of Frankfurt, Frankfurt, eRobert-Bosch Hospital Stuttgart, University of Tübingen, Tübingen, fMaria-Hilf Hospital Mönchengladbach, Mönchengladbach, and gMunicipal Hospital Krefeld, Krefeld, Germany 

Abstract

Background Although the use of premounted stents on a delivery balloon has almost completely eliminated the initially used hand-crimping procedure, no data are available that prove the superiority of one or the other approach on a randomized basis. Therefore, this study was designed to examine whether the use of premounted stents is comparable with the hand-crimping procedure. Methods A total of 123 patients (64 treated with unmounted stents, 59 treated with premounted stents) were examined in a multicenter, randomized, prospective study. There were no significant differences in patient characteristics between groups. Results Primary end points (acute, postinterventional [within 72 hours], and late complications related to the stenting procedure) were reached in 1 patient treated with an unmounted stent versus 2 patients with mounted stents (P = not significant). In patients with angiographic follow up (n = 84, mean follow-up period 6 ± 1 months), the total rate of restenosis was 27% (unmounted 12, mounted 11, P = not significant). Secondary end points were procedural success of stenting and maximal balloon inflation pressure needed for optimal stenting results by use of angiography. There were no differences in secondary end points for both techniques. The mean balloon pressure was 12.56 ± 2.1 atmospheres (unmounted) and 12.12 ± 1.92 atmospheres (mounted, P = not significant). Conclusion Stenting with premounted devices was demonstrated to have a similar clinical and angiographic outcome as the hand-crimping approach for maximal inflation pressure, procedural success, major cardiac events, and rate of restenosis after 6 months of follow up. Thus, the more convenient use of a premounted stent provides procedural safety and efficacy comparable with a hand-crimped system. (Am Heart J 2003;145:e4.)

Le texte complet de cet article est disponible en PDF.

Plan


 Supported by a grant of Guidant, Devices for Vascular Interventions.
☆☆ Reprint requests: Frank M. Baer, MD, Department III for Internal Medicine, University of Cologne, Joseph Stelzmannstr 9, 50924 Cologne, Germany.
 E-mail: frank.baer@uni-koeln.de
★★ 1097-6744/2003/$30.00 + 0


© 2003  Publié par Elsevier Masson SAS.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 145 - N° 2

P. K1-K6 - février 2003 Retour au numéro
Article précédent Article précédent
  • Reply
  • Jacopo Gianetti, Raffaele De Caterina
| Article suivant Article suivant
  • Low restenosis rate of the NIR coronary stent: Results of the Danish multicenter stent study (DANSTENT)—a randomized trial comparing a first-generation stent with a second-generation stent
  • Erik Jørgensen, Henning Kelbæk, Steffen Helqvist, Gunnar V.H. Jensen, Kari Saunamäki, Jens Kastrup, Jan Kyst Madsen, Lene Kløvgaard, Leif Thuesen, Anton Villadsen, Anton W.M. van Weert, Johan H.C. Reiber

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Elsevier s'engage à rendre ses eBooks accessibles et à se conformer aux lois applicables. Compte tenu de notre vaste bibliothèque de titres, il existe des cas où rendre un livre électronique entièrement accessible présente des défis uniques et l'inclusion de fonctionnalités complètes pourrait transformer sa nature au point de ne plus servir son objectif principal ou d'entraîner un fardeau disproportionné pour l'éditeur. Par conséquent, l'accessibilité de cet eBook peut être limitée. Voir plus

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2026 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.