S'abonner

Three-Dimensional Color Doppler Transesophageal Echocardiography for Mitral Paravalvular Leak Quantification and Evaluation of Percutaneous Closure Success - 04/11/14

Doi : 10.1016/j.echo.2014.08.019 
Eduardo Franco, MD a, , Carlos Almería, MD, PhD a, Jose Alberto de Agustín, MD, PhD a, Viviana Arreo del Val, MD b, José Juan Gómez de Diego, MD, PhD a, Miguel Ángel García Fernández, MD, PhD a, Carlos Macaya, MD, PhD a, Leopoldo Pérez de Isla, MD, PhD a, Eulogio Garcia, MD, PhD a
a Cardiovascular Institute, Hospital Clínico San Carlos, Madrid, Spain 
b Paediatric Cardiology Department, Hospital Universitario La Paz, Madrid, Spain 

Reprint requests: Eduardo Franco, MD, Cardiovascular Institute, Hospital Clínico San Carlos, Calle Profesor Martin Lagos s/n, 28040 Madrid, Spain.

Abstract

Background

Three-dimensional (3D) color Doppler transesophageal echocardiography (TEE) enables accurate planimetry of the effective regurgitant orifice (ERO) of a mitral paravalvular leak (PVL). The aim of this study was to evaluate the usefulness of this method to quantify paravalvular regurgitation and to assess percutaneous PVL closure success, compared with 3D planimetry of PVLs without using color-flow images (3D anatomic regurgitant orifice [ARO]).

Methods

Forty-six patients (59 mitral PVLs) who underwent 3D TEE to evaluate the indication of PVL closure procedure were retrospectively included. Receiver operating characteristic curves were compared to identify degree III and IV paravalvular regurgitation of 3D color ERO and 3D ARO measures. Forty patients underwent percutaneous PVL closure procedures; analysis was conducted to determine whether the undersizing of the closure devices according to 3D color ERO and 3D ARO measures was associated with PVL closure failure.

Results

Three-dimensional ERO measures showed better areas under the curve than 3D ARO measures and correlated better with the degree of paravalvular regurgitation. Three-dimensional color ERO major diameter ≥ 0.65 cm showed a positive predictive value of 87.1% and a negative predictive value of 94% to diagnose degree III and IV paravalvular regurgitation. For the 40 patients who underwent PVL closure procedures, the immediate technical success rate was 76.9%, and 1-year estimated survival was 69.5%. Closure device undersizing according to 3D color ERO length, but not other PVL measurements, was significantly associated with PVL closure failure (P = .007).

Conclusion

Three-dimensional ERO was superior to 3D ARO at identifying the presence of degree III and IV paravalvular regurgitation. The undersizing of closure devices according to 3D color ERO length was associated with failed closure procedures. Confirmatory prospective studies are encouraged.

Le texte complet de cet article est disponible en PDF.

Keywords : Paravalvular leak, Three-dimensional transesophageal echocardiography, Paravalvular mitral regurgitation, Effective regurgitant orifice, Percutaneous paravalvular leak closure procedure

Abbreviations : ARO, AUC, CI, ERO, PVL, ROC, 3D, 2D


Plan


© 2014  American Society of Echocardiography. Publié par Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 27 - N° 11

P. 1153-1163 - novembre 2014 Retour au numéro
Article précédent Article précédent
  • Comparison of Aortic Root Geometry with Bicuspid versus Tricuspid Aortic Valve: Real-Time Three-Dimensional Transesophageal Echocardiographic Study
  • Kentaro Shibayama, Kenji Harada, Javier Berdejo, Jun Tanaka, Hirotsugu Mihara, Yuji Itabashi, Takahiro Shiota
| Article suivant Article suivant
  • Impact of Implantable Transvenous Device Lead Location on Severity of Tricuspid Regurgitation
  • Karima Addetia, Francesco Maffessanti, Anuj Mediratta, Megan Yamat, Lynn Weinert, Joshua D. Moss, Hemal M. Nayak, Martin C. Burke, Amit R. Patel, Eric Kruse, Valluvan Jeevanandam, Victor Mor-Avi, Roberto M. Lang

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.