Microbiologic assessment of flexible gastrointestinal endoscope reprocessing using a pump-assisted sampling technique: an investigation involving all endoscopy units in Tianjin, China - 30/06/18
, Pei-Yong Ning, MB b, 1, Wei Zhang, MM b, Dong-Jing Yang, MM b, He Liu, MM a, Jun Liu, MM a, Jia Song, MB a, Chun-Nan Fei, MM aHighlights |
• | The failure rate of endoscope reprocessing in China is high. |
• | Pump-assisted sampling can improve the sensitivity of microbiologic cultures. |
• | The purer the final rinse water, the better. |
Abstract |
Background |
Microbiologic surveillance of flexible gastrointestinal endoscopes is recommended in several guidelines as the primary means of identifying reprocessing failures. This study aimed to evaluate the contamination level and prevalence of bacteria of post-reprocessing endoscopes and to access whether using a pump-assisted sampling method (PASM) improves the sensitivity of culture.
Methods |
All 59 endoscopy units in Tianjin, China, were investigated. The PASM and the conventional flushing sampling method (CFSM) were used to compare the results of the microbial culture. Logistic regression analysis was used to identify the influencing factors.
Results |
One hundred four (56.52%) flushing channel samples of gastrointestinal endoscopes were positive for culture, and the maximum bacterial concentration was 14,100 colony-forming units (CFU)/channel. One hundred fifty-one (82.07%) flushing samples were qualified according to the national standard of China (≤ 20 CFU/channel). The qualified rate of the samples collected by PASM was significantly lower than the qualified rate by CFSM (65.52% vs 89.68%). Using PASM (odds ratio [OR]: 4.257; 95% confidence interval [CI]: 1.870–9.690) would increase the sensitivity of culture. The use of purified water (OR: 0.288; 95% CI: 0.102–0.814) could reduce the risk of endoscope reprocessing failure.
Conclusion |
Many endoscopes fail to meet the national standard for microbial culture after reprocessing. Our results suggest that using a pump-assisted method could increase the sensitivity of the test.
Le texte complet de cet article est disponible en PDF.Key Words : Endoscope, Reprocessing, Pump-assisted sampling method, Contamination, Rinse water
Plan
| Funding/support: This study was supported by Financial Project Funding for the Tianjin Centers for Disease Control and Prevention in 2016- 2017. |
|
| Conflicts of interest: None declared. |
Vol 46 - N° 7
P. e43-e48 - juillet 2018 Retour au numéroBienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.
Déjà abonné à cette revue ?