EOS-based cup navigation: Randomised controlled trial in 78 total hip arthroplasties - 09/05/16

Doi : 10.1016/j.otsr.2016.02.006 
N. Verdier , A. Billaud, T. Masquefa, J. Pallaro, T. Fabre, C. Tournier
 Service de chirurgie orthopédique, hôpital Pellegrin, CHU de Bordeaux, place Amélie-Raba-Léon, 33000 Bordeaux, France 

Corresponding author. Tel.: +33 556 795 544; fax: +33 556 796 101.

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Abstract

Background

Minimising the risk of cup implantation outside the safe zone is among the objectives of navigation during total hip arthroplasty (THA). However, given the technical challenges raised by navigation when the patient is lying on the side, many surgeons still use the freehand technique. We conducted a randomised controlled trial to evaluate the new navigation system NAVEOS in the iliac plane, which is easily identified in the lateral decubitus position, with the objective of determining whether NAVEOS navigation decreased the frequency of cup implantation outside the safe zone compared to freehand cup positioning, without increasing the operative time or the frequency of complications.

Hypothesis

NAVEOS navigation decreases the frequency of cup positioning outside the safe zone compared to freehand positioning.

Material and methods

This randomised controlled trial compared cup positioning using NAVEOS navigation versus the freehand technique in patients undergoing primary THA. The safe zone was defined according to Lewinnek as 15±10° of radiological anteversion and 40±10° of radiological inclination. Cup position parameters were measured on computed tomography images obtained 3months after THA. The images were read by two independent observers who were blinded to group assignment. The primary evaluation criterion was cup position within the safe zone.

Results

A 1:1 randomisation scheme was used to assign 78 patients (mean age, 68years; age range, 44–91years) to NAVEOS navigation or freehand cup positioning. The two groups were comparable for age, gender distribution, body mass index, and preoperative functional scores. In the NAVEOS group, navigation was discontinued prematurely in 6 patients, because of technical difficulties (n=2) or a marked discrepancy with clinical findings (n=4); however, the intention-to-treat approach was used for the analysis. The proportion of cups in the safe zone was 67% (28/39) in the NAVEOS group and 38% (17/39) in the freehand group (P=0.012). Anteversion was within the 5–25° range for 72% (28/39) cups in the NAVEOS group and 46% (18/39) in the freehand group (P=0.021). Inclination was within the 30–50° range for 95% (37/39) of cups with NAVEOS navigation and 85% (33/39) with freehand positioning (P=0.135). The odds ratio for cup implantation outside the safe zone was significantly lower with NAVEOS compared to freehand positioning (0.54; 95% confidence interval, 0.31–0.91). Mean operative time was 74 (range, 45–115) minutes with NAVEOS navigation and 70 (range, 40–105) minutes with freehand positioning (P=0.382). Complications consisted of 1 case each of anterior dislocation and infection, both in the freehand group.

Discussion

Compared to freehand positioning, NAVEOS navigation significantly lowered the risk of cup positioning outside the safe zone, chiefly via improved achievement of the anteversion target. NAVEOS was not associated with increases in operative time or morbidity.

Level of evidence

II, randomised controlled trial with limited statistical power.

El texto completo de este artículo está disponible en PDF.

Keywords : Navigation, Hip arthroplasty, Computer-assisted surgery, Dislocation


Esquema


© 2016  Elsevier Masson SAS. Reservados todos los derechos.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 102 - N° 4

P. 417-421 - juin 2016 Regresar al número
Artículo precedente Artículo precedente
  • Contents
| Artículo siguiente Artículo siguiente
  • Reproducibility of digital measurements of lower-limb deformity on plain radiographs and agreement with CT measurements
  • G. Sorin, G. Pasquier, E. Drumez, A. Arnould, H. Migaud, S. Putman

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2024 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.