Modeling Cryotherapy Ice Ball Dimensions and Isotherms in a Novel Gel-based Model to Determine Optimal Cryo-needle Configurations and Settings for Potential Use in Clinical Practice - 21/04/16

Doi : 10.1016/j.urology.2016.02.012 
Taimur T. Shah a, b, c, * , Uri Arbel d, Sonja Foss e, Andrew Zachman e, Simon Rodney a, b, Hashim U. Ahmed a, b, Manit Arya a, b, f
a Division of Surgery and Interventional Science, University College London, London, UK 
b Department of Urology, University College London Hospitals NHS Foundation Trust, London, UK 
c Whittington Hospital, London, UK 
d Galil Medical Ltd., Yokneam, Israel 
e Galil Medical Inc., Arden Hills, MN 
f Princess Alexandra Hospital NHS Trust, Harlow, UK 

*Address correspondence to: Taimur T. Shah, M.B.B.S., B.Sc., M.R.C.S., Urology Research Group, 4th Floor, Room 4.23, 132 Hampstead Road, London NW1 2PS, UK.Urology Research Group4th Floor, Room 4.23, 132 Hampstead RoadLondonNW1 2PSUK

Abstract

Objective

To gain a better understanding of ice ball dimensions and temperature isotherms relevant for cell kill when using combinations of cryo-needles we set out to answer 4 questions: (1) what type of cryo-needle? (2) how many needles? (3) best spatial configuration? and (4) correct duty cycle percentage?

Methods

We conducted laboratory experiments to monitor ice ball dimensions and create multi-needle planar isotherm maps for 17G and 10G cryo-needles using a novel multi-needle thermocouple fixture within gel at body temperature. We tested configurations of 1-4 cryo-needles at duty cycles of 20%-100% with 1-2.5 cm spacing.

Results

Analysis of various combinations shows that a central core of ≤−40°C develops at a distance of ~1 cm around the cryo-needles. Temperature increases linearly from this point to the ice ball leading edge (0°C), which is a further ≈1 cm away. Thus, the −40°C isotherm is approximately 1 cm inside the leading edge of the ice ball.

The optimum distance between cryo-needles was 1.5-2 cm, at duty cycle settings of 70%-100%. At distances further apart or with lower duty cycle settings, ice balls either had a central core >−40°C or had an hourglass shape.

Conclusion

In answer to questions 1-3, tumor length, diameter, and shape will ultimately determine the number of needles and their configuration. However, we propose a conservative distance for cryo-needle placement between 1 and 1.5 cm should be adopted for clinical practice. In answer to question 4, using low duty cycle settings runs the risk of incomplete −40°C isotherm coverage of the tumor, and thus in routine practice we suggest that settings of 70%-100% are most appropriate.

Le texte complet de cet article est disponible en PDF.

Plan


 Financial Disclosure: Uri Arbel, Sonja Foss, and Andrew Zachman are employed by Galil Medical. The remaining authors declare that they have no relevant financial interests.
 Funding Support: Hashim U. Ahmed receives trial funding from Sonacare Inc, AMD SAS, Angiodynamics Inc, Sophiris Inc, and Trod Medical. He receives funding for lectures and attendance at conferences from Sonacare Inc, Astra Zeneca, Janssen, and Sanofi, as well as proctoring fees from Sonacare Inc and Nuada Medical Ltd. This study was run in conjunction with and funded by Galil Medical.


© 2016  The Authors. Publié par Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 91

P. 234-240 - mai 2016 Retour au numéro
Article précédent Article précédent
  • Author Reply
  • Yoshinobu Komai
| Article suivant Article suivant
  • PLK-1 Silencing in Bladder Cancer by siRNA Delivered With Exosomes
  • Kristin A. Greco, Carrie A. Franzen, Kimberly E. Foreman, Robert C. Flanigan, Paul C. Kuo, Gopal N. Gupta

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Elsevier s'engage à rendre ses eBooks accessibles et à se conformer aux lois applicables. Compte tenu de notre vaste bibliothèque de titres, il existe des cas où rendre un livre électronique entièrement accessible présente des défis uniques et l'inclusion de fonctionnalités complètes pourrait transformer sa nature au point de ne plus servir son objectif principal ou d'entraîner un fardeau disproportionné pour l'éditeur. Par conséquent, l'accessibilité de cet eBook peut être limitée. Voir plus

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2026 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.