S'abonner

Single-operator cholangioscopy and targeted biopsies in the diagnosis of indeterminate biliary strictures: a systematic review - 01/06/16

Doi : 10.1016/j.gie.2015.04.030 
Udayakumar Navaneethan, MD 1, , Muhammad K. Hasan, MD 1, Vennisvasanth Lourdusamy, MD 1, Basile Njei, MD, MPH 2, Shyam Varadarajulu, MD 1, Robert H. Hawes, MD 1
1 Center for Interventional Endoscopy, Institute for Minimally Invasive Surgery, Florida Hospital, Orlando, Florida, USA 
2 Department of Gastroenterology, Yale University, New Haven, Connecticut, USA 

Reprint requests: Udayakumar Navaneethan, MD, FACP, Center for Interventional Endoscopy, Institute for Minimally Invasive Surgery, Florida Hospital, 601 E Rollins Street, Orlando, FL 32814.

Abstract

Background and Aims

Evaluation of indeterminate biliary strictures by brush cytology and intraductal biopsies is limited by low sensitivity. The utility of SpyGlass peroral cholangioscopy for diagnosis of malignant biliary strictures and cholangiocarcinoma (CCA), in particular, remains unclear. Our aim was to study the utility of SpyGlass peroral cholangioscopy and targeted biopsy for diagnosis of malignant biliary strictures and CCA.

Methods

In this systematic review, PubMed and Embase databases were reviewed for studies published to October 2014. The main outcomes of interest were sensitivity, specificity, and diagnostic odds ratio (DOR) of SpyGlass cholangioscopy in the diagnosis of malignant biliary strictures.

Results

The search yielded 10 studies involving 456 patients. The pooled sensitivity and specificity of cholangioscopy-guided biopsies in the diagnosis of malignant biliary strictures was 60.1% (95% confidence interval [CI], 54.9%-65.2%) and 98.0% (95% CI, 96.0%-99.0%), respectively. The pooled DOR to detect malignant biliary strictures was 66.4 (95% CI, 32.1-137.5). Four studies included patients who had previous negative imaging and brushings and/or intraductal biopsies. Among these 4 studies, the pooled sensitivity and specificity for diagnosis of malignant biliary strictures was 74.7% (95% CI, 63.3%-84.0%) and 93.3% (95% CI, 85.1%-97.8%), respectively. The pooled DOR was 46.0 (95% CI, 15.4-138.1). Only 1 study directly compared the yield of SpyBite biopsies with standard brushings and biopsies. SpyBite biopsies had a sensitivity of 76.5% compared with brushings (5.8%) and biopsies (29.4%). Six studies specifically reported the role of cholangioscopy with targeted biopsies in the diagnosis of CCA. The pooled sensitivity and specificity to detect CCA was 66.2% (95% CI, 59.7%-72.3%) and 97.0% (95% CI, 94.0%-99.0%), respectively. The pooled DOR to detect CCA was 79.7 (95% CI, 32.7-194.7).

Conclusions

Our study suggests that SpyGlass cholangioscopy with SpyBite biopsies has moderate sensitivity for the diagnosis of malignant biliary strictures.

Le texte complet de cet article est disponible en PDF.

Abbreviations : CCA, DOR, FISH, LR, POC, QUADAS, SOPOC


Plan


 DISCLOSURE: U. Navaneethan is a consultant for AbbVie. R. Hawes and S. Varadarajulu are consultants for Olympus and Boston Scientific. All other authors disclosed no financial relationships relevant to this publication.
 If you would like to chat with an author of this article, you may contact Dr Navaneethan at udhaykumar81@gmail.com.


© 2015  American Society for Gastrointestinal Endoscopy. Publié par Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 82 - N° 4

P. 608 - octobre 2015 Retour au numéro
Article précédent Article précédent
  • The role of deep enteroscopy in the management of small-bowel disorders
  • ASGE Standards of Practice Committee, Mouen A. Khashab, Shabana F. Pasha, V. Raman Muthusamy, Ruben D. Acosta, David H. Bruining, Vinay Chandrasekhara, Krishnavel V. Chathadi, Mohamad A. Eloubeidi, Robert D. Fanelli, Ashley L. Faulx, Lisa Fonkalsrud, Suryakanth R. Gurudu, Loralee R. Kelsey, Shivangi Kothari, Jenifer R. Lightdale, John R. Saltzman, Aasma Shaukat, Amy Wang, Julie Yang, Brooks D. Cash, John M. DeWitt
| Article suivant Article suivant
  • Single-operator cholangioscopes in the diagnosis of cholangiocarcinoma: seeing is believing. Is belief enough to allow treatment?
  • Richard A. Kozarek

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Elsevier s'engage à rendre ses eBooks accessibles et à se conformer aux lois applicables. Compte tenu de notre vaste bibliothèque de titres, il existe des cas où rendre un livre électronique entièrement accessible présente des défis uniques et l'inclusion de fonctionnalités complètes pourrait transformer sa nature au point de ne plus servir son objectif principal ou d'entraîner un fardeau disproportionné pour l'éditeur. Par conséquent, l'accessibilité de cet eBook peut être limitée. Voir plus

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2026 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.