S'abonner

Calcification of Hydrophilic Acrylic Intraocular Lenses With a Hydrophobic Surface: Laboratory Analysis of 6 Cases - 28/07/16

Doi : 10.1016/j.ajo.2016.04.018 
Sotirios P. Gartaganis a, , Philipp Prahs b, Eftichia D. Lazari a, Panos S. Gartaganis a, Horst Helbig b, Petros G. Koutsoukos c
a Department of Ophthalmology, School of Medicine, University of Patras, Patras, Greece 
b Department of Ophthalmology, University Hospital of Regensburg, Regensburg, Germany 
c Department of Chemical Engineering, Laboratory of Inorganic and Analytical Chemistry, University of Patras, and FORTH-ICEHT, Patras, Greece 

Inquiries to Sotirios P. Gartaganis, Emeritus Professor of Ophthalmology, School of Medicine, University of Patras, 26504, Rion, Patras, GreeceEmeritus Professor of OphthalmologySchool of MedicineUniversity of Patras26504, RionPatrasGreece

Abstract

Purpose

To investigate the nature and characteristic features of deposits causing opacification of intraocular lenses (IOLs) based on the examination of clinical findings using scanning electron microscopy (SEM) and energy-dispersive x-ray spectroscopy (EDX) analysis.

Design

Retrospective, observational case series.

Methods

This is a multicenter study of 6 hydrophilic acrylic IOLs (Lentis LS-502-1; Oculentis GmbH, Berlin, Germany) with a hydrophobic surface that were explanted from 5 patients because of opacification. Three patients had an uncomplicated phacoemulsification. One patient underwent combined phacoemulsification and pars plana vitrectomy for retinal detachment and later silicone oil endotamponade owing to redetachment. The last patient had a pars plana vitrectomy and silicone oil instillation combined with phacoemulsification for tractive retinal detachment and diabetic retinopathy. The explanted lenses were submitted to our laboratory and were examined by SEM and EDX in order to identify the morphologic features and the composition of the deposits.

Results

SEM and EDX analyses confirmed the presence of calcific deposits in the interior of the opacified hydrophilic IOLs, with a pattern showing the formation of lumps on the surface. The lumps were due to subsurface formation of calcium phosphate crystalline deposits. The crystallite clusters seemed to diffuse from the IOL interior to the surface.

Conclusions

We demonstrated the calcification pattern of the hydrophilic IOL (Lentis LS-502-1) with a hydrophobic surface. Although hydrophilic acrylic lenses have a hydrophobic surface, the development of calcification is a possible threat initiating from the hydrophilic subsurface of the IOLs.

Le texte complet de cet article est disponible en PDF.

Plan


 Supplemental Material available at AJO.com.


© 2016  Elsevier Inc. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 168

P. 68-77 - août 2016 Retour au numéro
Article précédent Article précédent
  • Oral Bisphosphonates and Risk of Wet Age-Related Macular Degeneration
  • Zaid Mammo, Michael Guo, David Maberley, Joanne Matsubara, Mahyar Etminan
| Article suivant Article suivant
  • Predictors, Resource Utilization, and Short-term Costs of Laser Trabeculoplasty Versus Medication Management in Open-Angle Glaucoma
  • Neil M. Schultz, William B. Wong, Anne L. Coleman, Daniel C. Malone

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.