Ventricular fibrillation caused by electrocautery for pericardium hemostasis during thoracotomy - 14/12/17
Highlights |
• | A case of electrocautery-induced ventricular fibrillation in a patient undergoing elective thoracotomy was presented. |
• | A low-frequency leakage current from the electrocautery device was most likely responsible for the cardiac event. |
• | If bleeding is near the cardiac region, it is recommended to perform hemostasis using an ultrasonic scalpel. |
Vol 42