The effect of loop size on loop security and elongation of a knot - 30/01/20

Doi : 10.1016/j.otsr.2019.10.008 
Selim Ergün , Umut Akgün, Mustafa Karahan
 Acibadem Mehmet Ali Aydinlar University, School of Medicine, Department of Orthopedics and Traumatology, Istanbul, Turkey 

Corresponding author at: Acibadem Kozyatagı Hastanesi, Okur sokak, no: 24, Kozyatağı, 34734, Istanbul.Acibadem Kozyatagı HastanesiOkur sokak, no: 24Kozyatağı34734Istanbul

Bienvenue sur EM-consulte, la référence des professionnels de santé.
Article gratuit.

Connectez-vous pour en bénéficier!

Abstract

Introduction

While repairing a teared rotator cuff tendon with suture anchors and horizontal mattress suture configurations, knots should be secure at time zero while approximating the tendon to the bone, otherwise any failure in loop security may cause undesired clinical results. Optimum distance between suture limbs passed through the tendon, in other words the bite size, is still not clear in the literature. The aim of this study was to test the effect of loop size, which is directly related to the bite size, on loop security and elongation of a knot.

Hypothesis

We hypothesized that a knot with a smaller loop size would be more secure. We asked if a knot with shorter circumference (1) would offer a better knot security; (2) would produce less elongation following repeated traction cycles.

Material and methods

Two parallel metal rods in 3.0mm diameter were fixed to load cells of dynamic testing machine. Four groups, from A to D, had the initial rod to rod distances of 2-4-6-8mm respectively (n=10). Surgeon's knots were prepared with 2/0 Ultrabraid® sutures around the rods. A tension meter was used for tying each half hitch under equal tension. Crosshead distances were recorded after 7N pre-load and subsequent 1000 repetitive cyclic loads between 7-30N.

Results

Elongations after a 7N preload for groups A to D were 0.107mm (±0.006), 0.143mm (±0.018), 0,16mm (±0.025), 0.185mm (±0.018) respectively. This increase was significant (p<0.05, power>0.95) between each group except between groups B and C. Maximum elongations after 1000th cycle for groups A to D were 0.32mm (±0.124), 1.12mm (±0.333), 1.162mm (±0.211), 1.292mm (±0.241) respectively. Only samples in group A (0.732mm±0.124) elongated significantly less than others (p<0.05, power >0.95). No knots unravelled or ruptured.

Discussion

This study basically reports that a knot with a shorter loop circumference has superior properties regarding loop security and resistance to elongation. From the perspective of clinical importance, shorter distance between suture limbs of mattress configuration may provide a more secure fixation of the rotator cuff tendon to the bone.

Level of evidence

II.

Le texte complet de cet article est disponible en PDF.

Keywords : Loop circumference, Mattress configuration, Loop security, Maximal elongation


Plan


© 2019  Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 106 - N° 1

P. 35-38 - février 2020 Retour au numéro
Article précédent Article précédent
  • Assessment of the Quality and Reliability of the Information on Rotator Cuff Repair on YouTube
  • Haluk Celik, Omer Polat, Cagri Ozcan, Savas Camur, Bekir Eray Kilinc, Metin Uzun
| Article suivant Article suivant
  • 3D motion analysis of latissimus dorsi tendon transfer in patients with posterosuperior rotator cuff tears: Analysis of proprioception and the ability to perform ADLS
  • Pit Hetto, Sarah Erhard, Mirjam Thielen, Sebastian I. Wolf, Felix Zeifang, Stefan van Drongelen, Michael W. Maier

Bienvenue sur EM-consulte, la référence des professionnels de santé.

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.