S'abonner

Failure to eradicate non-tuberculous mycobacteria upon disinfection of heater–cooler units: results of a microbiological investigation in northwestern Italy - 31/10/20

Doi : 10.1016/j.jhin.2020.08.023 
S. Ditommaso , M. Giacomuzzi, G. Memoli, C.M. Zotti
 Department of Public Health and Pediatrics, University of Turin, Italy 

Corresponding author. Address: University of Turin, Public Health and Microbiology Department, Via Santena 5 bis, Turin 10126, Italy. Tel.: +39 (0)116705841; fax: +39 (0)116705881.University of TurinPublic Health and Microbiology DepartmentVia Santena 5 bisTurin10126Italy

Summary

Background

Heater–cooler units (HCUs) used during cardiopulmonary bypass may become colonized with non-tuberculous mycobacteria (NTM), including Mycobacterium chimaera. Recently, a worldwide investigation conducted in hospitalized infected patients has detected M. chimaera in several Stockert 3T HCUs manufactured by LivaNova.

Aim

Microbiological surveillance on Stockert 3T (LivaNova) and Maquet HCU40 (Getinge) devices as well as an evaluation of the efficacy of their recommended decontamination protocols.

Methods

A total of 308 water samples were collected from 29 HCUs: 264 samples were collected from 17 Stockert 3T HCUs and 44 samples from 12 Maquet HCU40 devices. Samples were tested for total viable counts (TVCs) at both 22 and 36°C, Pseudomonas aeruginosa, coliform bacteria, and NTM. The microbiological surveillance began in June 2017 and ran until October 2019.

Findings

A total of 308 HCU water samples were analysed, 65.5% of which yielded NTM. The most frequently colonized device with NTM was the Stockert 3T (88.2%), with a frequency of positive samples of 59.5% (157/264). The Maquet HCU40 devices less frequently yielded NTM (33.3%), with a frequency of positive water samples of 13.6% (6/44). Disinfection procedures were effective in reducing TVCs of bacteria with the exception of NTM species. NTM were detected in both pre-disinfection (50.1%) and post-disinfection (55.7%) samples, and no significant association was found between disinfection and NTM results both in Stockert 3T and Maquet HCU40 devices.

Conclusion

This study suggests that manufacturers' procedures for disinfection are ineffective and/or inadequate. Until effective disinfection protocols become available, the only way to minimize the risk of NTM contamination is to closely monitor the water quality in the HCU, keep it as clean as possible, and treat it like any other biohazardous material.

Le texte complet de cet article est disponible en PDF.

Keywords : Mycobacterium chimaera, Polymerase chain reaction, Heater–cooler unit, Cardiac surgery, Microbiological monitoring, Disinfection


Plan


© 2020  The Healthcare Infection Society. Publié par Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 106 - N° 3

P. 585-593 - novembre 2020 Retour au numéro
Article précédent Article précédent
  • The use of germicidal ultraviolet light, vaporized hydrogen peroxide and dry heat to decontaminate face masks and filtering respirators contaminated with a SARS-CoV-2 surrogate virus
  • L.F. Ludwig-Begall, C. Wielick, L. Dams, H. Nauwynck, P-F. Demeuldre, A. Napp, J. Laperre, E. Haubruge, E. Thiry
| Article suivant Article suivant
  • Susceptibility of patients receiving chemotherapy for haematological malignancies to scabies
  • H. Hosoi, S. Nishikawa, Y. Kida, T. Kishi, S. Murata, M. Iwamoto, Y. Toyoda, Y. Yamada, T. Ikeda, T. Sonoki

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Elsevier s'engage à rendre ses eBooks accessibles et à se conformer aux lois applicables. Compte tenu de notre vaste bibliothèque de titres, il existe des cas où rendre un livre électronique entièrement accessible présente des défis uniques et l'inclusion de fonctionnalités complètes pourrait transformer sa nature au point de ne plus servir son objectif principal ou d'entraîner un fardeau disproportionné pour l'éditeur. Par conséquent, l'accessibilité de cet eBook peut être limitée. Voir plus

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2026 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.