Comparison of Methods for Measuring Cyclodeviation - 14/04/21

Doi : 10.1016/j.ajo.2020.11.005 
Laura Liebermann a, Sarah R. Hatt a, David A. Leske a, Lindsay D. Klaehn a, Andrea M. Kramer a, Jonathan M. Holmes a, b,
a Department of Ophthalmology, Mayo Clinic, Rochester, Minnesota, USA 
b Department of Ophthalmology and Vision Science, University of Arizona-Tucson, Tucson, Arizona, USA 

Inquiries to Jonathan M. Holmes, Department of Ophthalmology and Vision Science, University of Arizona-Tucson, 655 N. Alvernon Way, Suite 204, Tucson, Arizona 85711, USADepartment of Ophthalmology and Vision ScienceUniversity of Arizona-Tucson655 N. Alvernon Way, Suite 204TucsonArizona85711USA

Abstract

Purpose

To compare the double-Maddox rod test with other methods of measuring cyclodeviation

Design

Retrospective cohort study.

Methods

We retrospectively identified 153 adults in a clinical practice with cyclodeviation assessed using double-Maddox rods, of whom 105 were also assessed using fusible synoptophore targets, 73 using nonfusible synoptophore targets, 118 using single-Maddox rod, and 43 using fundus photography. Relationships between double-Maddox rod and other tests were evaluated by calculating mean differences with 95% confidence intervals (CI), intraclass correlation coefficients (ICC), and Bland-Altman plots with linear regression.

Results

Synoptophore cross-in-circle targets and the largest (of right or left) single-Maddox rod values were similar to double-Maddox values (mean differences: −1.2° and 0.1°, respectively; ICC: 0.79 and 0.82, respectively). Synoptophore house targets measured less excyclodeviation (mean difference: −2.7°; ICC: 0.71). Mean summed single-Maddox rod values were somewhat similar to double-Maddox values (mean difference: 1.5°; ICC: 0.85), but differences increased with greater cyclodeviation (r2 = 0.2678; P < .001). Fundus photographs showed large, uncorrelated differences compared with double-Maddox rod test, when summing right and left eyes and when using the largest of right or left (mean differences: 12.2° and 6.2°; ICC: −0.02 and 0.21, respectively), and differences increased with greater cyclodeviation (r2 = 0.4094; P < .001 and r2 = .1143; P= .03, respectively).

Conclusions

There was good agreement between double-Maddox and the largest single- Maddox test values and synoptophore cross-in-circle targets but poorer agreement with other tests. Further study is needed to understand which measurements best reflect true cyclodeviation and relationships with symptoms.

Le texte complet de cet article est disponible en PDF.

Plan


© 2020  Elsevier Inc. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 224

P. 332-342 - avril 2021 Retour au numéro
Article précédent Article précédent
  • Correlations Between Choriocapillaris and Choroidal Measurements and the Growth of Geographic Atrophy Using Swept Source OCT Imaging
  • Yingying Shi, Qinqin Zhang, Hao Zhou, Liang Wang, Zhongdi Chu, Xiaoshuang Jiang, Mengxi Shen, Marie Thulliez, Cancan Lyu, William Feuer, Luis de Sisternes, Mary K. Durbin, Giovanni Gregori, Ruikang K. Wang, Philip J. Rosenfeld
| Article suivant Article suivant
  • Comment on: Allogenic Simple Limbal Epithelial Transplantation Versus Amniotic Membrane Grafting in the Early Management of Severe-Grade Ocular Chemical Injuries—A Retrospective Comparative Study
  • Pedro-Ivan Navarro, Alejandro Lichtinger, José Bareño

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Elsevier s'engage à rendre ses eBooks accessibles et à se conformer aux lois applicables. Compte tenu de notre vaste bibliothèque de titres, il existe des cas où rendre un livre électronique entièrement accessible présente des défis uniques et l'inclusion de fonctionnalités complètes pourrait transformer sa nature au point de ne plus servir son objectif principal ou d'entraîner un fardeau disproportionné pour l'éditeur. Par conséquent, l'accessibilité de cet eBook peut être limitée. Voir plus

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2026 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.