Impact of dose reduction and the use of an advanced model-based iterative reconstruction algorithm on spectral performance of a dual-source CT system: A task-based image quality assessment - 26/06/21

Doi : 10.1016/j.diii.2021.03.002 
Joël Greffier , Djamel Dabli, Aymeric Hamard, Philippe Akessoul, Asmaa Belaouni, Jean-Paul Beregi, Julien Frandon
 Department of Medical Imaging, CHU Nimes, Univ Montpellier, Medical Imaging Group Nimes, EA 2992, Bd Pr Robert Debré, 30029 Nîmes Cedex 9, France 

Corresponding author.

Bienvenue sur EM-consulte, la référence des professionnels de santé.
Article gratuit.

Connectez-vous pour en bénéficier!

Highlights

The use of low energy levels on virtual monoenergetic images (VMI) generated by dual-source CT improves the low contrast resolution but alters noise texture and increases noise magnitude.
The use of ADMIRE reduces the noise magnitude for low energy levels on VMIs obtained with dual-source CT.
ADMIRE increases the detectability of enhanced iodine lesions for low keV on VMI obtained with dual-source CT.

Le texte complet de cet article est disponible en PDF.

Abstract

Purpose

To assess the impact of dose reduction and the use of an advanced modeled iterative reconstruction algorithm (ADMIRE) on image quality in low-energy monochromatic images from a dual-source dual energy computed tomography CT (DSCT) platform.

Materials and methods

Acquisitions on an image-quality phantom were performed using DSCT equipment with 100/Sn150 kVp for four dose levels (CTDIvol: 20/11/8/5mGy). Raw data were reconstructed for six energy levels (40/50/60/70/80/100 keV) using filtered back projection and two levels of ADMIRE (A3/A5). Noise power spectrum (NPS) and task-based transfer function (TTF) were calculated on virtual monoenergetic images (VMIs). Detectability index (d′) was computed to model the detection task of two enhanced iodine lesions as function of keV.

Results

Noise-magnitude was significantly reduced between 40 to 70 keV by −56±0% (SD) (range: −56%–−55%) with FBP; −56±0% (SD) (−56%–−56%) with A3; and −57±1% (SD) (range: −57%–−56%) with A5. The average spatial frequency of the NPS peaked at 70 keV and decreased as ADMIRE level increased. TTF values at 50% were greatest at 40 keV and shifted towards lower frequencies as the keV increased. The detectability of both lesions increased with increasing dose level and ADMIRE level. For the simulated lesion with iodine at 2mg/mL, d’ values peaked at 70 keV for all reconstruction types, except for A3 at 20 mGy and A5 at 11 and 20 mGy, where d’ peaked at 60 keV. For the other simulated lesion, d’ values were highest at 40 keV and decreased beyond.

Conclusion

At low keV on VMIs, this study confirms that iterative reconstruction reduces the noise magnitude, improves the spatial resolution and increases the detectability of enhanced iodine lesions.

Le texte complet de cet article est disponible en PDF.

Keywords : Multidetector computed tomography, Dual energy, Iterative reconstruction, Task-based image quality assessment, Phantom study

Abbreviations : ADMIRE, CT, CTDIvol, DECT, FBP, HU, IR, MBIR, NPWE, NPS, ROI, SECT, SD, TTF, VMI


Plan


© 2021  Société française de radiologie. Publié par Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 102 - N° 7-8

P. 405-412 - juillet 2021 Retour au numéro
Article précédent Article précédent
  • From ovary to brain: Specific imaging features of ovarian teratomas associated with anti-N-methyl-d-aspartate receptor encephalitis
  • Stephanie Nougaret, Teresa Margarida Cunha
| Article suivant Article suivant
  • A fast and reproducible method to estimate left atrial volume using cardiac computed tomography
  • Francesca Pitocco, Gilles Soulat, Imen Menif, Etienne Charpentier, Kim-Diep D. Tran, Elie Mousseaux

Bienvenue sur EM-consulte, la référence des professionnels de santé.

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.