Quantitative and qualitative optimisation of dosimetry in computed tomography explorations of the temporal bone using two iterative reconstruction algorithms - 28/10/21

Doi : 10.1016/j.neuri.2021.100017 
Olivier Legeas a, David Bourhis b, Marc Garetier c, Adele Pennaneach d, Philippe Meriot a, Julien Ognard a, e, f, , Douraied Ben Salem a, e, f
a Imaging Unit, University Hospital of Brest, Bd Tanguy Prigent, 29609 Brest Cedex, France 
b Radiation Physics and Protection Unit, University Hospital of Brest, Bd Tanguy Prigent, 29609 Brest Cedex, France 
c Imaging Unit, Military Teaching Hospital of Brest, Rue Colonel Fonferrier, 29240 Brest Cedex, France 
d Radiology Unit, Cornouaille Hospital of Quimper, 14Bis Avenue Yves Thépot, 29107 Quimper cedex, France 
e Laboratoire de traitement de l'information médicale—LaTIM (Inserm UMR 1101), 22, avenue Camille Desmoulins, CS 93837-29238 Brest Cedex 3, France 
f Clinical Forensic Medicine and Pathology, University Hospital of Brest, Bd Tanguy Prigent, 29609 Brest Cedex, France 

Corresponding author at: Medical Imaging Service, University Hospital of Cavale Blanche, Bd Tanguy Prigent, 29609 Brest Cedex, France.Medical Imaging ServiceUniversity Hospital of Cavale BlancheBd Tanguy PrigentBrest Cedex29609France

Bienvenue sur EM-consulte, la référence des professionnels de santé.
Article gratuit.

Connectez-vous pour en bénéficier!

Highlights

Analysis of Iterative Reconstruction algorithms on phantoms help to maximise dose reduction while maintaining a desired level of readability.
We measured in an image quality phantom the evolution of the Contrast-to-Noise Ratio for each IR algorithm depending on the CDTIvol.
Acquisition parameters were established involving the anthropomorphic phantom for an evaluation of the image quality at each iteration degree.
The optimal use were: Siemens: SAFIRE 3 with a CDTIvol with a dose reduction of 65%; GE: ASiR 20% with a dose reduction of 20%.

Le texte complet de cet article est disponible en PDF.

Abstract

Purpose: Iterative reconstruction (IR) has recently demonstrated its ability to reduce the dose of X-rays used in the exploration of several anatomical regions. However, there is no study involving a qualitative and quantitative comparison of the various IRs available for exploring the temporal bone. The aim of this study is to analyse two IR algorithms (ASiR®, statistical iterative reconstruction and SAFIRE®, model based iterative reconstruction). Methods: Using a Catphan® phantom, we studied the quantitative relationship of the contrast on noise/CDTIvol/IR. We verified the absence of a reduction in spatial resolution during the use of IRs by measuring MTF50%. Based on these data we acquired a series of images on an anthropomorphic phantom with an authentic human temporal bone for each level of iteration; the purpose of this was to evaluate them qualitatively. The CDTIvol of temporal bone IRs was adjusted based on the measurements taken with the Catphan phantom in order to acquire images with a constant CNR. Results: The resulting measurements demonstrated a reduction in noise without a loss of spatial resolution in the two manufacturers' systems. The quantitative study by four radiologists allowed us to optimise CDTIvol and the level of iteration. This optimisation made it possible to reduce the dose in the two manufacturers' systems, and the most effective of these algorithms enabled a reduction in CDTIvol of 64.6% during explorations of the temporal bone. Conclusion: The two analysed iterative reconstruction algorithms responded well to the objectives of dose reduction in the exploration of the temporal bone.

Le texte complet de cet article est disponible en PDF.

Keywords : Iterative reconstruction, Dosimetry, ASiR, SAFIRE, Temporal bone


Plan


 This article has been handled by another Editor, Jean-Christophe Gentric during the whole editorial process.


© 2021  The Author(s). Publié par Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 1 - N° 4

Article 100017- décembre 2021 Retour au numéro
Article précédent Article précédent
  • Contents
| Article suivant Article suivant
  • Quantitative evaluation of gray matter alterations in patients with mesial temporal lobe epilepsy (MTLE)
  • Zhensheng Li, Quwen Gao, Kairun Peng, Jian Lin, Wei Wang, Weimin Wang, Bingmei Deng

Bienvenue sur EM-consulte, la référence des professionnels de santé.

Elsevier s'engage à rendre ses eBooks accessibles et à se conformer aux lois applicables. Compte tenu de notre vaste bibliothèque de titres, il existe des cas où rendre un livre électronique entièrement accessible présente des défis uniques et l'inclusion de fonctionnalités complètes pourrait transformer sa nature au point de ne plus servir son objectif principal ou d'entraîner un fardeau disproportionné pour l'éditeur. Par conséquent, l'accessibilité de cet eBook peut être limitée. Voir plus

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2026 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.