S'abonner

Repeatability of the Accommodative Response Measured by the Grand Seiko Autorefractor in Children With and Without Amblyopia and Adults - 23/03/22

Doi : 10.1016/j.ajo.2021.10.019 
Ryan N. Chinn 1, Aparna Raghuram 1, Molly K. Curtiss 2, Alyssa M. Gehring 2, Ana Juric De Paula 2, Tawna L. Roberts 2, 3, #, #
1 Department of Ophthalmology, Boston Children's Hospital/Harvard Medical School, Boston, Massachusettes (R.N.C., A.R.) 
2 Akron Children's Hospital, Akron, Ohio (M.C., A.G., A.J.D.P., T.L.R.) 
3 Spencer Center for Vision Research, Byers Eye Institute at Stanford University, Palo Alto, California (T.L.R.) 

Inquiries to Tawna L. Roberts, OD, PhD, 2370 Watson Ct, Suite 100, Palo Alto, CA 94303. P: 650-498-3154.2370 Watson Ct, Suite 100Palo AltoCA94303

Résumé

PURPOSE

To assess test-retest repeatability of the accommodative response (AR) in children with and without amblyopia and adults using the Grand Seiko autorefractor.

DESIGN

Prospective reliability assessment.

METHODS

Test-retest of accommodation was obtained while participants viewed 20/150 sized letters at 33 cm using the Grand Seiko autorefractor in children 5 to <11 years with amblyopia (n=24) and without amblyopia (n=36), and adults 18 to <35 years (n=34). Bland-Altman 95% limits of agreement (LOA) and intraclass correlation coefficients (ICCs) were used to assess repeatability and reliability. The AR between the fellow and amblyopic eyes of children with amblyopia and eye 1 and eye 2 of the visually normal participants was assessed using group comparisons.

RESULTS

The 95% LOA of the AR was greatest in the amblyopic eyes (–1.25 diopters [D], 1.62 D) of children with amblyopia. The 95% LOA were similar between the fellow eyes (–0.88 D, 0.74 D) of children with amblyopia and both eyes of the children without amblyopia (eye 1: –0.68 D, 0.71 D; eye 2: –0.59 D, 0.70 D) and the adults (eye 1: 95% LOA = –0.49 D, 0.45 D; eye 2: LOA = –0.66 D, 0.67 D). ICCs revealed the Grand Seiko autorefractor as a reliable instrument for measuring AR.

CONCLUSIONS

The Grand Seiko autorefractor was more repeatable and reliable when measuring the AR in children and adults without amblyopia than in the amblyopic eye in children with amblyopia. It is recommended that multiple measures of the AR be obtained in amblyopic eyes to improve the precision of measures.

Le texte complet de cet article est disponible en PDF.

Plan


 Supplemental Material available at AJO.com.


© 2021  Elsevier Inc. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 236

P. 221-231 - avril 2022 Retour au numéro
Article précédent Article précédent
  • Longitudinal Assessment of Ellipsoid Zone Recovery Using En Face Optical Coherence Tomography After Retinal Detachment Repair
  • Natalia Figueiredo, David Sarraf, Frederic Gunnemann, Srinivas R. Sadda, Aditya Bansal, Alan R. Berger, David T. Wong, Radha P. Kohly, Peter J. Kertes, Roxane J. Hillier, Rajeev H. Muni
| Article suivant Article suivant
  • Nonmydriatic Ultra-Widefield Fundus Photography in a Hematology Clinic Shows Utility for Screening of Sickle Cell Retinopathy
  • Ishrat Ahmed, Tejus Pradeep, Morton F. Goldberg, T.Y. Alvin Liu, Ahimsa Aradhya, Manuela P. Montana, Natalie Photiadis, Elizabeth Williams, Brantley Smith, Jing Tian, Sophie M. Lanzkron, Adrienne W. Scott

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Elsevier s'engage à rendre ses eBooks accessibles et à se conformer aux lois applicables. Compte tenu de notre vaste bibliothèque de titres, il existe des cas où rendre un livre électronique entièrement accessible présente des défis uniques et l'inclusion de fonctionnalités complètes pourrait transformer sa nature au point de ne plus servir son objectif principal ou d'entraîner un fardeau disproportionné pour l'éditeur. Par conséquent, l'accessibilité de cet eBook peut être limitée. Voir plus

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2026 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.