Abstract

Objectives

Prosthetic socket is the contact interface between the stump and the prosthesis, and also the interface component that transmits forces from the stump to the prosthesis distal. The current prosthetic socket fit is a major factor affecting rehabilitation, especially with the stump volume fluctuations. The main goal of this article is to design an adjustable frame-type prosthetic socket with constant force to adapt to the stump volume fluctuations.

Materials and methods

In this paper, an adjustable frame-type prosthetic socket with constant force is designed. The constant force device is designed based on the superelasticity of the shape memory alloy for maintaining constant stump-socket interface stress and automatically adapting to certain volume fluctuations. The constant force extrusion performance of this prosthetic socket was verified and optimized by finite element analysis.

Results

The results suggest that the constant force unit may maintain constant interface stress. According to the optimization results, the shape memory alloy dimensional parameters could be selected according to different requirements.

Conclusion

The adjustable frame-type prosthetic socket allows the user to adjust the socket volume through the cable system and has a large heat dissipation area. The constant force unit maintains constant interface stress and automatically adapts to stump volume fluctuations.

Le texte complet de cet article est disponible en PDF.

Graphical abstract

Le texte complet de cet article est disponible en PDF.

Highlights

Volume fluctuations cause interface pressure mismatch.
Constant force extrusion for stump may solve interface pressure mismatch.
An adjustable frame-type socket with constant force is designed.
Constant force extrusion of the prosthetic socket is verified by FEA.

Le texte complet de cet article est disponible en PDF.

Keywords : Interface stress, Prosthetic socket, Shape memory alloy, Stump, Superelasticity, Volume fluctuation


Plan


© 2022  AGBM. Publié par Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 44 - N° 1

Article 100731- février 2023 Retour au numéro
Article précédent Article précédent
  • Automatic Detection of Severely and Mildly Infected COVID-19 Patients with Supervised Machine Learning Models
  • M.T. Huyut
| Article suivant Article suivant
  • CRISPR-OTE: Prediction of CRISPR On-Target Efficiency Based on Multi-Dimensional Feature Fusion
  • J. Xie, M. Liu, L. Zhou

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.