Évaluation de la performance clinique et analytique du système de mesure de glycémie Vital Check® MM-1200/bandelettes MS-2 dans la vraie vie - 11/10/23

Evaluation of the clinical and analytical performance in real life of the blood glucose monitoring system Vital Check® MM-1200/MS-2 strips

Doi : 10.1016/j.mmm.2023.04.002 
Amar Tebaibia 1, , Rachid Malek 2, Lakhdar Griene 3, Ali El Madhi Haddam 4, Nassim Nouri 5, Anwar Cherrak 6, Krim Belkacem 7, Samira Abrouk 8, Mourad Semrouni 9, Kamel Gasmi 10, Said Gari 10
1 Université Alger 1, établissement public hospitalier El Biar (Ex. Birtraria), service de médecine interne, El Biar, Alger, Algérie 
2 Université de Sétif, CHU de Sétif, service de médecine interne-diabétologie, Sétif, Algérie 
3 Université Alger 1, laboratoire d’hormonologie - centre Pierre et Marie Curie, Alger, Algérie 
4 Université Alger 1, CHU Bab El Oued, service de diabétologie-endocrinologie, Alger, Algérie 
5 Université de Constantine, CHU de Constantine, service d’endocrinologie-diabétologie, Constantine, Algérie 
6 Université d’Oran, CHU d’Oran, service de médecine interne, Oran, Algérie 
7 Maison du diabétique El Harrach, Alger, Algérie 
8 Méthodologie et Bio statistique consultante, Alger, Algérie 
9 Université Alger 1, CHU Beni-Messous, service de diabétologie-endocrinologie, Alger, Algérie 
10 Vital Care, production distribution et logistique, Alger, Algérie 

Amar Tebaibia, Établissement public hospitalier El Biar, service de médecine interne, 126 avenue Mustapha Ali Khodja, El Biar, Alger, Algérie.Établissement public hospitalier El Biar, service de médecine interne, 126 avenue Mustapha Ali KhodjaEl BiarAlgerAlgérie

Résumé

Introduction

La bonne gestion du diabète, les décisions thérapeutiques et la sécurité des patients dépendent de la performance analytique et clinique du système de mesure des glycémies capillaires. Celles-ci, effectuées par les patients, présentent des écarts par rapport aux glycémies veineuses mesurées au laboratoire.

Méthodes et objectifs

Étude nationale (Algérie), prospective, multicentrique, dans la vraie vie. L’objectif de l’étude était d’évaluer, dans la vraie vie, la concordance des résultats entre les glycémies capillaires mesurées par le système Vital Check® MM-1200 et les glycémies veineuses à jeun mesurées par un automate de laboratoire de biologie, chez des adultes atteints de diabète de type 1 (DT1) ou de diabète de type 2 (DT2). Les critères de l’exactitude analytique et de la performance clinique de la norme ISO 15197:2013 ont été évalués chez 462 patients DT1 ou DT2 issus de six centres.

Résultats

La concordance analytique globale était de 79,9 % [intervalle de confiance à 95 % : 79,9 ± 3,6] avec un coefficient de variation (CV) de 5,8 % et un biais moyen de -1,5 %. La performance clinique, en utilisant la grille d’erreur de Parkes, a objectivé que 99.2 % des valeurs étaient dans les zones A et B, pour lesquelles les différences entre glycémies capillaires et veineuses ne sont pas considérées comme déterminantes. Les facteurs limitant la performance de la mesure du glucose capillaire par le système Vital Check® MM-1200, sont liés aux patients, à l’environnement et à l’usage correct du système par les patients.

Conclusion

La concordance analytique du système de mesure de glycémie capillaire était sous-optimale. Elle était liée à des facteurs d’interférences. L’éducation thérapeutique du patient, pour la bonne pratique de l’autosurveillance glycémique (ASG), est essentielle pour limiter l’influence des facteurs exogènes sur les valeurs de glycémie capillaire mesurées qui pourraient avoir un impact négatif sur la décision clinique et thérapeutique.

Le texte complet de cet article est disponible en PDF.

Summary

Introduction

Successful diabetes management, treatment decisions, and patient safety depend on the analytical and clinical performance of capillary glucose monitoring systems. These, carried out by patients, show deviations from venous blood glucose levels measured in the laboratory.

Methods and aims

A national (Algerian), real-life, prospective, multicentric study. The objective of our study was to evaluate, in real life, the concordance between capillary glycemia measured by Vital Check® MM-1200 and fasting venous glycemia measured in the laboratory, in adults with type 1 or type 2 diabetes. The analytical accuracy criteria and the ISO 15197:2013 clinical performance were assessed in 462 adults with type 1 or type 2 diabetes, cared for in 6 centers.

Results

The overall analytical concordance was 79.9% [95% CI: 79.9 ± 3.6] with a coefficient of variation of 5.8% and a mean bias of -1.5%. The clinical performance, using the Parkes error grid, indicated that 99.2% of the values were in zones A and B, where differences between capillary and venous glycemia are not serious. Factors limiting the performance of the method of measuring capillary glucose were related to the patients, the environment and the proper use of this monitoring system.

Conclusion

Analytical concordance of this capillary blood glucose measurement system was suboptimal because of interfering factors. Therapeutic education, for the good practice of self-monitoring of blood glucose (SMBG), is essential to limit exogenous factors that influence the measured values of capillary glucose, that could have a negative impact on clinical and therapeutic decisions.

Le texte complet de cet article est disponible en PDF.

Mots clés : Autosurveillance glycémique, Vital Check® MM-1200, Concordance analytique, Performance clinique, Éducation thérapeutique, Algérie

Keywords : Self-monitoring of blood glucose, Vital Check® MM-1200, Analytical concordance, Clinical performance, Patient education, Algeria


Plan


© 2023  Publié par Elsevier Masson SAS.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 17 - N° 6

P. 507-516 - octobre 2023 Retour au numéro
Article précédent Article précédent
  • Brèves
| Article suivant Article suivant
  • Impact de la mise en place d’une prise en charge psycho-nutritionnelle sur la qualité de vie des patients obèses
  • Elise Le Roy Feret, Estelle Caron, Jérémy Hermann, Anne-Laure Proust, Camille Gaulier, Erika Cornu, Anne-Cécile Paepegaey

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Elsevier s'engage à rendre ses eBooks accessibles et à se conformer aux lois applicables. Compte tenu de notre vaste bibliothèque de titres, il existe des cas où rendre un livre électronique entièrement accessible présente des défis uniques et l'inclusion de fonctionnalités complètes pourrait transformer sa nature au point de ne plus servir son objectif principal ou d'entraîner un fardeau disproportionné pour l'éditeur. Par conséquent, l'accessibilité de cet eBook peut être limitée. Voir plus

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2025 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.