S'abonner

Recurrent Valve Frame Infolding in the 34mm Self-Expanding Medtronic Evolut Pro+ Transcatheter Valve System: A Single Centre Experience - 16/12/23

Doi : 10.1016/j.hlc.2023.08.015 
Karan Rao, MD, MPH a, b, c, Alexandra Baer, BSN, MSHM b, Peter Hansen, MBBS, PhD a, b, Ravinay Bhindi, MBBS, PhD a, b, c, ⁎
a Department of Cardiology, Royal North Shore Hospital, Sydney, NSW, Australia 
b Department of Cardiology, North Shore Private Hospital, Sydney, NSW, Australia 
c Faculty of Medicine and Health, The University of Sydney, Sydney, NSW, Australia 

∗Corresponding author at: Department of Cardiology, Royal North Shore Hospital, Reserve Road, St. Leonards, NSW 2065, AustraliaDepartment of CardiologyRoyal North Shore HospitalReserve RoadSt. LeonardsNSW2065Australia

Abstract

Background

Valve frame infolding (VFI) is a rare complication (1-3%) of transcatheter aortic valve implantation (TAVI), which primarily occurs in large sized self-expanding valves. We report the incidence of VFI in a small group of patients with highly calcified and extra-large aortic annuli, who underwent implantation of the newer generation 34mm Medtronic Evolut Pro+ (EP+) valve system.

Methods

A retrospective review was conducted on all patients presenting to a single centre experienced with TAVI in Sydney, NSW, Australia, between June and October 2022, for transfemoral TAVI using the Medtronic 34mm EP+ valve system. VFI was diagnosed using fluoroscopy, and pre-procedure computed tomography (CT) was analysed offline.

Results

VFI occurred in four of 10 patients who underwent TAVI using the 34mm EP+ system. Between VFI and non-VFI patients, the annular size, aortic angulation and total calcium volume was similar. Calcium distribution between the coronary cusps was symmetrical in non-VFI patients (37%, 33% and 30%), compared to 52% in the non-coronary cusp (NCC) in VFI patients. The mean ellipticity index was higher in VFI versus non-VFI patients (22% vs 14%).

Conclusion

The 34mm Medtronic EP+ valve system is vital for safe and effective treatment of a niche subgroup of patients with extra-large annuli and extensive calcification, however operators should be aware of the increased incidence of VFI. Heavy eccentric calcification at the NCC, as well as using the cusp-overlap implant technique may exacerbate VFI, while multiplanar fluoroscopic screening may improve recognition.

Le texte complet de cet article est disponible en PDF.

Keywords : TAVI, Aortic stenosis


Plan


© 2023  Australian and New Zealand Society of Cardiac and Thoracic Surgeons (ANZSCTS) and the Cardiac Society of Australia and New Zealand (CSANZ). Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 32 - N° 12

P. 1482-1488 - décembre 2023 Retour au numéro
Article précédent Article précédent
  • Variability in Contemporary Heparin Prescription and Activated Clotting Time Monitoring During Percutaneous Coronary Intervention: Call for Up-To-Date Evidence-Based Guidelines
  • Omar Al-Mukhtar, Dion Stub, Christopher M. Reid, Sidney Lo, Jeffrey Lefkovits, Antony Walton, Derek P. Chew, Andy Yong, Stephen J. Nicholls, Nicholas Cox, Karlheinz Peter, William Chan
| Article suivant Article suivant
  • Low Transvalvular Flow Rate in Patients Undergoing Transcatheter Aortic Valve Implantation (TAVI) Is a Predictor of Mortality: The TFR-TAVI Study
  • Ayman Al-atta, Mohamed Farag, Visvech Jeyalan, Amjad Gazzal Asswad, Alexandra Thompson, Tim Irvine, Richard Edwards, Rajiv Das, Azfar Zaman, Mohammad Alkhalil

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Elsevier s'engage à rendre ses eBooks accessibles et à se conformer aux lois applicables. Compte tenu de notre vaste bibliothèque de titres, il existe des cas où rendre un livre électronique entièrement accessible présente des défis uniques et l'inclusion de fonctionnalités complètes pourrait transformer sa nature au point de ne plus servir son objectif principal ou d'entraîner un fardeau disproportionné pour l'éditeur. Par conséquent, l'accessibilité de cet eBook peut être limitée. Voir plus

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2026 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.