A multicenter study on the effects of different methods of disinfecting medical external-use ultrasound probes - 25/01/24

Doi : 10.1016/j.ajic.2023.06.007 
Lingling Chen, MSc a, b, c, Hongmei Zhang, MSc a, b, c, Fan Chang, MSc d, Hua Yu, BSc d, Man Lu, MD e, Jing Zhao, BSc f, Xiaojuan Ma, MD g, Hong Zhou, BSc h, Liuying Zhou, MSc i, Jing Tan, BSc j, Jingyu Wang, BSc k, Dingqiang Luo, BSc l, Min Chen, BSc m, Lixue Yin, MSc a, b, c,
a Department of Cardiovascular Ultrasound and Non-invasive Cardiology, Sichuan Academy of Medical Sciences & Sichuan Provincial People's Hospital, Chengdu, Sichuan, China 
b Ultrasound in Cardiac Electrophysiology and Biomechanics Key Laboratory of Sichuan Province, Sichuan Provincial People's Hospital, University of Electronic Science and Technology of China, Chengdu, China 
c Ultrasound Medical Quality Control Center of Sichuan Province, Chengdu, Sichuan, China 
d Department of Clinical Laboratory Medicine, Sichuan Academy of Medical Sciences & Sichuan Province People’s Hospital, Chengdu, Sichuan, China 
e Department of Ultrasound, Sichuan Cancer Hospital, Chengdu, Sichuan, China 
f Department of Ultrasound, Sichuan Provincial Maternity and Child Health Care Hospital, Chengdu, Sichuan, China 
g Department of Ultrasound, Chengdu First People’s Hospital, Chengdu, Sichuan, China 
h Department of Ultrasound, Chengdu Third People’s Hospital, Chengdu, Sichuan, China 
i Department of Ultrasound, Chengdu Women’s and Children’s Central Hospital, Chengdu, Sichuan, China 
j Department of Ultrasound, the People's Hospital of Wenjiang District, Chengdu, Sichuan, China 
k Department of Ultrasound, the First People’s Hospital of Longquanyi District, Chengdu, Sichuan, China 
l Department of Ultrasound, Sichuan Tianfu New Area People’s Hospital, Chengdu, Sichuan, China 
m Department of Ultrasound, Sichuan Modern Hospital, Chengdu, Sichuan, China 

Address correspondence to Lixue Yin, Department of Cardiovascular Ultrasound and Non-invasive Cardiology, Sichuan Academy of Medical Sciences & Sichuan Provincial People's Hospital, 32# W. Sec 2, 1st Ring Rd. Qing Yang District, Chengdu, Sichuan 610041, China.Department of Cardiovascular Ultrasound and Non-invasive Cardiology, Sichuan Academy of Medical Sciences & Sichuan Provincial People's Hospital32# W. Sec 2, 1st Ring Rd. Qing Yang DistrictChengduSichuan610041China

Résumé

Background

Microbial contamination of external-use ultrasound probes is a serious and overlooked issue. We assessed the effects of different methods of disinfecting medical external-use ultrasound probes.

Methods

On-site disinfection experiments were conducted at 10 hospitals; the tips and sides of external-use ultrasound probes were sampled before and after disinfection, and 3 disinfection methods were assessed (use of a new ultraviolet [UV] ultrasound probe disinfector, wiping with ordinary paper towels or cleaning with disinfectant wipes).

Results

For the new UV probe disinfector, the median microbial death rates for the tips and sides of the external-use ultrasound probe were 93.67% and 97.50%, respectively, which were higher than those for wiping with paper towels and cleaning with disinfectant wipes (12.50% and 10.00% and 20.00% and 21.42%, respectively); the rates of microorganisms exceeding the standard were 15.0% and 13.3%, respectively, which were lower than those for wiping with paper towels and cleaning with disinfectant wipes (53.3% and 60.0% and 46.7% and 38.3%, respectively). The death rates of different species of microorganisms were high, ranging from 87.5% to 100%.

Conclusions

The new UV ultrasound probe disinfector significantly reduced the risk of potential nosocomial infections according to the low microbial death rate for conventional disinfection methods.

Le texte complet de cet article est disponible en PDF.

Highlights

Microbial contamination of external-use ultrasound probes is serious and overlooked.
Wiping with paper towels and cleaning with disinfectant wipes are traditional disinfection methods.
The new ultraviolet ultrasound probe disinfector can destroy the DNA or RNA of pathogens on the probe and block their replication and reproduction and have higher disinfection efficiency than the above disinfection methods.

Le texte complet de cet article est disponible en PDF.

Key Words : Bacterial contamination, Disinfection


Plan


 Funding/support: Hunan Health Medical Technology Co., Ltd, the manufacturer of the new-type UV ultrasonic probe disinfector, funded the microbiological test of samples, but did not participate in the result analysis or the writing of any part of the manuscript.
 Conflicts of interest: None to report.


© 2023  Association for Professionals in Infection Control and Epidemiology, Inc.. Publié par Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 52 - N° 2

P. 167-175 - février 2024 Retour au numéro
Article précédent Article précédent
  • Evaluation of the effects of wiping decontamination for filter cartridges of elastomeric half-mask respirators (EHMRs)
  • Evanly Vo, Matthew Horvatin, Ziqing Zhuang, Caitlin McClain, Rebecca Streeter, Jeremy Brannen, Nichole L. Suhon
| Article suivant Article suivant
  • Personal stethoscope disinfection practices and bacterial contamination: A cross-sectional study at the University Hospital Emergency Department in Belgrade, Serbia
  • Ana Jovanovic, Katarina Paunovic, Marko Ercegovac, Dusan Popovic, Dragana Davidovic

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Elsevier s'engage à rendre ses eBooks accessibles et à se conformer aux lois applicables. Compte tenu de notre vaste bibliothèque de titres, il existe des cas où rendre un livre électronique entièrement accessible présente des défis uniques et l'inclusion de fonctionnalités complètes pourrait transformer sa nature au point de ne plus servir son objectif principal ou d'entraîner un fardeau disproportionné pour l'éditeur. Par conséquent, l'accessibilité de cet eBook peut être limitée. Voir plus

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2026 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.