Energy consumption in MRI: Determinants and management options - 24/02/24

Doi : 10.1016/j.neurad.2023.12.001 
Mateusz Chodorowski a, , Julien Ognard a, b, Àlex Rovira c, Jean-Christophe Gentric a, d, David Bourhis d, e, Douraied Ben Salem a, b
a Service d'Imagerie Médicale, CHU Brest, Univ. Brest, Boulevard Tanguy Prigent, Brest CEDEX, 29609, France 
b INSERM UMR 1101, Laboratoire de Traitement de L'Information Médicale - LaTIM, Université de Bretagne Occidentale, 22, Avenue C. Desmoulins, Brest 29238 Cedex 3, France 
c Section of Neuroradiology, Department of Radiology, Hospital Universitari Vall d'Hebron, Universitat Autònoma de Barcelona, Spain 
d Inserm, UMR 1304 (GETBO), Western Brittany Thrombosis Study Group, Université de Bretagne Occidentale, Brest, France 
e Service de Physique Médicale, CHU Brest, France 

Corresponding author.

Abstract

Background

Energy consumption awareness is a known concern, and radiology departments have energy-intensive consuming machines. The means of energy consumption management in MRI scanners have yet to be evaluated.

Purpose

To measure the MRI energy consumption and to evaluate the means to reduce it.

Materials and methods

Data was retrieved for two MRI scanners through the hospital's automated energy consumption measurement software. After correlation with picture archiving and communication system (PACS) files, they were segmented by machine and mode (as follows: stand-by, idle and active) and analyzed. Active mode data for a specific brain MRI protocol have been isolated, and equivalent low energy consuming protocol was made. Both were performed on phantom and compared. Same protocol was performed on a phantom using 3.0T 16 and 32 head channels coils. Multiples sequences were acquired on phantom to evaluate power consumption.

Results

Stand-by mode accounted for 60 % of machine time and 40 % of energy consumption, active mode accounted for 20 % machine time and 40 % energy consumption, idle mode for 20 % imachine time and 20 % consumption. DWI and TOF sequences were the most consuming in our brain-MRI protocol. The low energy consuming protocol allowed a saving of approximately 10 % of energy consumption, which amounted for 0.20€ for each examination. This difference was mainly due to an energy consumption reduction of the DWI sequence. There were no difference in consumption between a 3.0T 16 and 32 channels head coils. Sequence's active power and duration (especially considering slice thickness) have to be taken into account when trying to optimize energy consumption.

Conclusion

There are two key factors to consider when trying to reduce MRI scan energy consumption. Stand-by mode energy consumption has to be taken into account when choosing an MRI scan, as it can't be changed further on. Active mode energy consumption is dependent of the MRI protocols used, and can be reduced with sequences adaptation, which must take into account sequence's active power and duration, on top of image quality.

Le texte complet de cet article est disponible en PDF.

Keywords : MRI, Energy consumption, Green radiology, Stand-by mode

Abbreviations : MRI, DWI, TOF, SWIp, SWAN, PACS, SAR, EPI


Plan


© 2023  Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 51 - N° 2

P. 182-189 - mars 2024 Retour au numéro
Article précédent Article précédent
  • Is non-contrast MRI sufficient to detect meningioma residue after surgery?
  • S. Motillon Alonso, F. Lersy, F.D. Ardellier, H. Cebula, F. Proust, A. Onofrei, A. Chammas, S. Kremer
| Article suivant Article suivant
  • Extended treatment in cerebral ischemia score 2c or 3 as goal of successful endovascular treatment is associated with clinical benefit
  • Adrian Karlsson, Katarina Jood, Isabella M. Björkman-Burtscher, Alexandros Rentzos

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.