Quantification of Straylight Induced by Silicone Oil Adherent to Intraocular Lenses of Different Materials - 16/05/24

Doi : 10.1016/j.ajo.2023.11.018 
Maximilian Hammer 1, 2, Leoni Britz 1, 2, Sonja Schickhardt 2, Ingo Lieberwirth 3, Donald Munro 2, Philipp Uhl 4, Alexander Scheuerle 1, 2, Ramin Khoramnia 1, Grzegorz Łabuz 1, 2, Gerd Uwe Auffarth 1, 2,
1 From the Department of Ophthalmology, University Clinic Heidelberg (M.H., L.B., A.S., R.K., G.Ł., G.U.A.) 
2 David J Apple Laboratory for Vision Research (M.H., L.B., S.S., D.M., A.S., R.K., G.Ł., G.U.A.) 
3 Heidelberg, Department of Physical Chemistry of Polymers, Max Planck Institute for Polymer Research, Mainz (I.L.) 
4 Institute for Pharmacy and Molecular Biotechnology, Heidelberg (P.U.), Germany. 

Inquiries to Gerd Uwe Auffarth, David J Apple Laboratory for Vision Research, Heidelberg, GermanyDavid J Apple Laboratory for Vision ResearchHeidelbergGermany

Résumé

Purpose

A complication of using silicone oil as an intraocular endotamponade is its adhesion to intraocular lenses (IOLs). Forward light scattering is a measure to quantify the optical disturbance caused by adherent oil droplets. We tested the straylight caused by silicone oil adhesion to different IOLs and examined whether an approved cleaning solution, F4H5, reverses the induced straylight.

Design

An experimental study.

Methods

Two hydrophobic acrylic IOL models and 1 hydrophilic model with a hydrophobic surface (n = 8 per model: 24 lenses) had straylight measured before contact with silicone oils, providing a baseline for subsequent testing: 12 lenses with lighter-than-water silicone oil (Siluron 2000) and 12 with heavier-than-water oil (Densiron 68). The final measurement was performed after cleansing with F4H5 when we used scanning electron and light microscopy to detect surface changes.

Results

Straylight was majorly increased in IOLs with adherent silicone oil (baseline vs adherent oil median 3.1 [2.1, 3.9] and 39.7 [22.7, 87.8] deg2/sr, respectively; P < .001). No difference was seen between heavier- and lighter-than-water silicone oils. Between IOL types, induced straylight varied significantly, with 1 hydrophobic model reaching the highest average straylight. F4H5 significantly reduced straylight values in all IOL types (median 9.4 [5.4, 13.8] deg2/sr). The microscopy revealed surface changes on the IOLs even after cleaning.

Conclusions

Silicone oil adhesion to IOLs can induce amounts of straylight known to cause severe optical disturbance. F4H5 cleansing solution reversed straylight values to only slightly increased values. We found no difference in straylight formation between the lighter- and heavier-than-water silicone oils.

Le texte complet de cet article est disponible en PDF.

Plan


 Supplemental Material available at AJO.com.


© 2023  The Author(s). Publié par Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 262

P. 192-198 - juin 2024 Retour au numéro
Article précédent Article précédent
  • Outcomes of Strabismus Surgery Following Teprotumumab Therapy
  • Grant Hilliard, Jaron Pruett, Sean P. Donahue, Federico G. Velez, Jason H. Peragallo, Lauren C. Ditta, Mehdi Tavakoli, Mary E. Hoehn, Annie F. Kuo, Maanasa Indaram, Natalie C. Kerr
| Article suivant Article suivant
  • Binocular Home Treatment for Amblyopia: Gains Stable for One Year
  • Tamara Wygnanski-Jaffe, Avital Moshkovitz, Burton J. Kushner, Michael Belkin, Oren Yehezkel, CureSight Pivotal Trial Group, Abraham Spierer, Tamara Wygnanski-Jaffe, Nethanel Zitzer, Dan Cohen, Ahuva Shpigelman, Maoz Hadash, Ilya Ortenberg, Rinat Cohen, Hana Leib, Majd Arow, Reut Parness, Luba Rodov, Alexandra Goz, Haia Katz, Anabel Bazov, Chaim Nissen, Gabriel Avraham, Emad Borsha, Idit Keynann, Tali Aviv, Nathalie Corcos, Keren Roll, Eedy Mezer, Vered Brucker, Meital Abecassis, Ronen Rabinovich, Eran Laster, Ronit Politi, Hila Givoni, Ahed Amitirat, Chiya Robert Barrett, Adelina Zioni, Katty Kuperman, Yael Crocos

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Elsevier s'engage à rendre ses eBooks accessibles et à se conformer aux lois applicables. Compte tenu de notre vaste bibliothèque de titres, il existe des cas où rendre un livre électronique entièrement accessible présente des défis uniques et l'inclusion de fonctionnalités complètes pourrait transformer sa nature au point de ne plus servir son objectif principal ou d'entraîner un fardeau disproportionné pour l'éditeur. Par conséquent, l'accessibilité de cet eBook peut être limitée. Voir plus

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2026 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.