L’utilisation de la caméra thermique smartphone compatible (FLIR One) pour le repérage préopératoire des perforantes de DIEP - 22/12/24

Using a smartphone-compatible thermal camera (FLIR One) for preoperative mapping of DIEP perforators

Doi : 10.1016/j.anplas.2024.09.002 
J. Roukoz a, , G. Bilger a, M. Chatard a, F. Hollard-Kissel a, M. Gédor b, M. Brix a, c, E. Simon a, c
a Service de chirurgie maxillofaciale et plastique, CHU de Nancy, hôpital Central, 29, avenue du Maréchal-de-Lattre-de-Tassigny-CO60034, 54035 Nancy cedex, France 
b Service de santé publique, CHR de Mercy, 1, allée du Château, 57085 Ars-Laquenexy, France 
c Faculté de médecine, université de Lorraine, 9, avenue de la Forêt-de-Haye, 54500 Vandœuvre-lès-Nancy, France 

Auteur correspondant.

Résumé

Introduction

La réussite des chirurgies impliquant des lambeaux libres dépend notamment d’une cartographie préopératoire précise des artères perforantes. Différentes techniques d’imagerie, comme le Doppler et l’angiographie tomodensitométrique, sont généralement employées, chacune ayant ses avantages et inconvénients en termes de coût, de précision et de risques pour le patient. L’objectif principal de notre étude est de comparer l’efficacité du dispositif FLIR One pour la détection de ces artères, en comparaison avec les méthodes traditionnelles. L’imagerie thermique, semblant pouvoir être une alternative plus simple, moins coûteuse et moins invasive lors planification chirurgicale.

Matériel et méthode

L’étude, réalisée au centre hospitalier régional de Nancy, a inclus 25 lambeaux libres (DIEP) sur une cohorte exclusivement féminine de 22 patientes, avec un suivi allant de 2022 à 2023. Avant l’intervention, un angioscanner abdominopelvien était effectué, suivi d’une évaluation complémentaire avec l’imagerie thermique FLIR One et un examen Doppler acoustique la veille de l’opération. Cette démarche visait à fournir une cartographie précise des perforantes vasculaires pour chaque patient. Trois opérateurs différents ont réalisé ces marquages, et les résultats ont ensuite été comparés avec les observations peropératoires. Le protocole d’imagerie incluait également un examen Doppler pour valider les résultats de l’imagerie thermique. Des analyses statistiques avec indices de corrélation intraclasse (ICC) ont été effectuées pour évaluer la corrélation entre les différentes méthodes de localisation préopératoire des vaisseaux perforants.

Résultats

Vingt-deux patientes ayant bénéficié 25 lambeaux de DIEP pour des reconstructions mammaires, trois techniques d’imagerie ont été utilisées pour identifier les perforantes vasculaires : l’imagerie thermique FLIR, le Doppler acoustique et l’angioscanner. FLIR a identifié le plus grand nombre de perforantes (n=137), suivi du Doppler acoustique (n=128) et de l’angioscanner (n=126). Lors de la comparaison avec les résultats peropératoires, 66 % des perforantes identifiées par FLIR ont été confirmées, 70 % pour le Doppler acoustique, et 95 % pour l’angioscanner. L’indice de corrélation intraclasse (ICC) montre une corrélation significative entre ces techniques d’imagerie et les résultats peropératoires. Le FLIR a démontré une forte corrélation avec les observations peropératoires (ICC de 0,74, p<0,001), suivie d’une corrélation modérée avec le Doppler acoustique (ICC de 0,56, p<0,03) et l’angioscanner (ICC de 0,52, p<0,006).

Conclusion

L’étude conclut que l’imagerie thermique FLIR est un outil à la fois fiable et efficace pour la localisation des perforantes vasculaires. Bien que l’étude et le FLIR aient leurs propres limitations, l’outil présente plusieurs avantages comme la facilité d’utilisation, la rapidité et le coût abordable. Ces caractéristiques rendent le FLIR particulièrement attractif comme complément aux méthodes traditionnelles de détection, telles que le Doppler acoustique et la tomodensitométrie (TDM).

Le texte complet de cet article est disponible en PDF.

Summary

Introduction

The success of surgeries involving free flaps largely depends on accurate preoperative mapping of perforator arteries. Various imaging techniques, such as Doppler ultrasound and CT angiography, are typically used, each having its advantages and disadvantages in terms of cost, accuracy, and patient risk. The main objective of our study is to compare the effectiveness of the FLIR One device for detecting these arteries compared with traditional methods. Thermal imaging appears to be a simpler, less expensive, and less invasive alternative for surgical planning.

Material and method

The study, conducted at the Regional Hospital Center of Nancy, included 25 free flaps (DIEP) on an exclusively female cohort of 22 patients, with follow-up from 2022 to 2023. Before the procedure, an abdominal-pelvic CT angiography was performed, followed by additional evaluation with thermal imaging using FLIR One and acoustic Doppler examination on the eve of the operation. This approach aimed to provide precise mapping of vascular perforators for each patient. Three different operators performed these markings, and the results were then compared with intraoperative observations. The imaging protocol also included a Doppler examination to validate the results of the thermal imaging. Statistical analyses with intraclass correlation coefficients (ICC) were performed to evaluate the correlation between different preoperative localization methods of perforating vessels.

Results

In 22 patients undergoing 25 DIEP flaps for breast reconstructions, three imaging techniques were used to identify vascular perforators: thermal imaging with FLIR, acoustic Doppler, and CT angiography. FLIR identified the most perforators (n=137), followed by acoustic Doppler (n=128) and CT angiography (n=126). Comparing these with intraoperative results, 66% of perforators identified by FLIR were confirmed, 70% for acoustic Doppler, and 95% for CT angiography. The ICCs shows a significant correlation between these imaging techniques and intraoperative results. FLIR demonstrated a strong correlation with intraoperative observations (ICC of 0.74, P<0.001), followed by a moderate correlation with acoustic Doppler (ICC of 0.56, P<0.03) and CT angiography (ICC of 0.52, P<0.006).

Conclusion

The study concludes that thermal imaging with FLIR is a reliable and effective tool for locating vascular perforators. Although the study and FLIR have their own limitations, the tool presents several advantages such as ease of use, speed, and affordability. These characteristics make FLIR particularly attractive as a complement to traditional detection methods, such as acoustic Doppler and CT angiography.

Le texte complet de cet article est disponible en PDF.

Mots clés : Traitement des images médicales, Lambeaux perforants, Reconstruction mammaire, Repérage

Keywords : Medical image processings, Perforator flap, Reconstruction, Breast


Plan


© 2024  Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 70 - N° 1

P. 10-18 - janvier 2025 Retour au numéro
Article précédent Article précédent
  • Stratégie de suivi des patients dans un centre SOS main universitaire labellisé FESUM-Intégration des étudiants de 3e cycle : une étude prospective
  • F. Delas, S. Tomczak, CM. Garcia-Doldan, D. Coquerel-Beghin, R. Beccari, O. Dhellemmes, I. Auquit-Auckbur
| Article suivant Article suivant
  • Submental liposuction under local anesthesia – About 695 patients
  • M. Aldihani, N. Assaf, S. Garson, R. Sinna

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Elsevier s'engage à rendre ses eBooks accessibles et à se conformer aux lois applicables. Compte tenu de notre vaste bibliothèque de titres, il existe des cas où rendre un livre électronique entièrement accessible présente des défis uniques et l'inclusion de fonctionnalités complètes pourrait transformer sa nature au point de ne plus servir son objectif principal ou d'entraîner un fardeau disproportionné pour l'éditeur. Par conséquent, l'accessibilité de cet eBook peut être limitée. Voir plus

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2026 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.