Preventing microbial contamination during radiological imaging: Experimental evaluation of a multiuse contrast media infusion system incorporating sequential one-way valves within a dual-safety design - 13/08/25

Doi : 10.1016/j.ajic.2025.06.006 
Matheus (Roland) van Rens, RN, MaANP a, , Kevin Hugill, RN, BSc, MSc, PhD b, L. Natalia Galvan, BS c, Benjamin Tanner, PhD c
a Radboudumc University, Nijmegen, The Netherlands 
b Independent Scholar, Cumbria, UK 
c Microchem Laboratory, Round Rock, TX, USA 

Address correspondence to Matheus (Roland) van Rens, RN, MaANP, Radboudumc University, Postbus 9102, 6500 HC Nijmegen, The Netherlands.Radboudumc UniversityPostbus 9102Nijmegen6500 HCThe Netherlands

Résumé

Background

Health care-associated infections are a significant cause of patient morbidity, mortality, and health care expenditure. In diagnostic imaging, multiuse contrast media infusion systems are increasingly common; however, their use raises concerns regarding cross-contamination risk.

Methods

A controlled experimental model was used to assess the ability of a multiuse infusion system to prevent microbial contamination during simulated clinical conditions. Escherichia coli and MS2 bacteriophage were selected to model bacterial and viral contamination risks, respectively. Inocula were introduced at key connection points during two 12-hour experimental sessions. The primary outcome was the presence or absence of microbial growth in retrieved fluid samples. Control testing was conducted to validate sterility, microbial viability, and experimental integrity.

Results

Positive and negative control testing performed as expected. No microbial growth was detected in any of the 51 experimental fluid samples. This corresponds to a 95% confidence upper bound of 5.8% for undetected contamination events.

Conclusions

The system's design, including sequential dual 1-way valves effectively prevented bacterial and viral contamination under laboratory conditions. These findings support its potential for safe multiuse in radiological settings.

Le texte complet de cet article est disponible en PDF.

Highlights

Dual-valve contrast infusion system prevented contamination in simulated tests.
No bacterial or viral growth in 51 uses; 95% CI upper contamination bound: 5.8%.
Viral challenge using a small surrogate virus exceeds current FDA test standards.
Multiuse infusion in radiology is safe and sustainable with no infection risk.
Device design is crucial: dual-valve safety zone blocks contamination reliably.

Le texte complet de cet article est disponible en PDF.

Key Words : Patient safety, Infection prevention, Microbial contamination testing, Multiuse medical devices, MRI/CT contrast media injector systems


Plan


 Funding/support: This study was supported by P&R Medical BV, who provided funding for laboratory testing. To maintain objectivity, the experimental protocol was independently designed by the study team, and all laboratory work was conducted by third-party scientists at an accredited facility. Data analysis and manuscript preparation were performed without sponsor involvement. The authors affirm that the funder had no role in data interpretation or reporting.
 Conflicts of interest: None to report.
 Ethics approval: This was a laboratory bench study that did not involve human or animal participants. Ethical approval was therefore not required.


© 2025  The Authors. Publié par Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 53 - N° 9

P. 962-966 - septembre 2025 Retour au numéro
Article précédent Article précédent
  • Impact of empowerment of onco-hematological patients on the incidence of catheter-related bloodstream infections
  • Rodrigo Reghini, Ivan Leonardo Avelino França e Silva, Silvia Figueiredo Costa, Thaís Guimarães
| Article suivant Article suivant
  • Provincial PPE Safety Coach Program evaluation in continuing care settings in Alberta
  • Kaethel Decker, Heather Gagnon, Abram Gutscher, Meaghan Brierley, Sara Mallinson, Jenine Leal

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Elsevier s'engage à rendre ses eBooks accessibles et à se conformer aux lois applicables. Compte tenu de notre vaste bibliothèque de titres, il existe des cas où rendre un livre électronique entièrement accessible présente des défis uniques et l'inclusion de fonctionnalités complètes pourrait transformer sa nature au point de ne plus servir son objectif principal ou d'entraîner un fardeau disproportionné pour l'éditeur. Par conséquent, l'accessibilité de cet eBook peut être limitée. Voir plus

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2026 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.