Automatic continuous P0.1 measurements during weaning from mechanical ventilation: a clinical study - 08/01/26
, Maamar Adel 1, 2, Guillot Pauline 1, 2, Quelven Quentin 1, 2, Coirier Valentin 1, 2, Painvin Benoit 1, 2, Tadie Jean-Marc 1, 2, 3, Terzi Nicolas 1, 2, 3, Gacouin Arnaud 1, 2, 3Abstract |
Background |
In critically ill patients, weaning from mechanical ventilation (MV) includes spontaneous breathing trial (SBT) usually followed by a reventilation period in order to recover from the alveolar derecruitement induced by the SBT. The measurement of occlusion pressure during the first 100 ms of an airway occlusion (P 0.1 ) one of the non-invasive tools available for estimating the respiratory drive, is a determinant of patient respiratory effort. This clinical study explores the use of non-invasive continuous monitoring of occlusion pressure automatically calculated by ventilators in the first 100 ms of airway occlusion (P 0.1 vent ) during SBT and reventilation periods. The study aimed to investigate patient or respirator factors influencing P 0.1 vent as well as the association of P 0.1 vent values with extubation success or failure.
Patients and Methods |
This prospective observational study, conducted from February 2022 to April 2023, included adult patients intubated for more than 24 h and screened for extubation weaning. SBTs were performed for one hour with zero pressure support and zero end-expiratory pressure (PS0 ZEEP). Reventilation followed for an hour with pressure support (8–12 cmH 2 O) and PEEP (5 cmH 2 O). Data included patient characteristics, ventilator parameters and extubation outcomes.
Results |
The study involved 224 measurements from 212 patients, with 157 successful extubations, 46 extubation failures at day 7 and 21 SBT failures. P 0.1 vent mean values were significantly higher for extubation failures and SBT failures compared to successful extubations ( p < 0.001). Delta P 0.1 vent ((P 0.1 vent reventilation - P 0.1 vent SBT)/ P 0.1 vent SBT) was significantly different according to whether extubation was a success or a failure: 0.21 (0.02–0.62) cm H 2 O vs. P 0.1 vent vs. 1.12 (0.54–2.38) cm H 2 O; p < 0.0001 respectively. Values significantly differed in both the SBT and the reventilation periods whether or not patients had previous ARDS: 1.08 (0.70; 2.02) cmH 2 O vs. 0.80 (0.54; 1.28) cmH 2 O respectively ( p = 0.003). Noteworthy, P 0.1 vent values were influenced by airway humidification systems (0.92 (0.57; 1.54) cmH 2 O with humidification vs. 1.27 (0.91; 2.24) cmH 2 O without, p = 0.003).
Conclusion |
The delta of P 0.1 vent values between SBT and reventilation are higher for patients who fail extubation, especially for those who had ARDS. While elevated P 0.1 vent values were associated with extubation failure, the overlap in values limits its usefulness as a reliable predictor.
Le texte complet de cet article est disponible en PDF.Keywords : P 0.1 automatic measurement , Mechanical ventilation weaning, Ventilator accuracy, Spontaneous breathing trial
Keywords : Medical and Health Sciences, Clinical Sciences