Capteur ECG intelligent pour la synchronisation des séquences en IRM et le monitorage des patients - 31/05/11

Smart MR ECG sensor for sequence synchronization and patient monitoring

Doi : 10.1016/j.irbm.2011.01.010 
J. Felblinger a, b, , d , J.-P. Blondé e, S. Jovanovic a, b, L. Rousselet a, b, J. Oster a, b, C. Pasquier a, b, d, J. Pascal e, f, L. Zhou e, f, N. Pillet e, f, J.-B. Schell e, f, M. Ayachi e, f, P.-Y. Marie c, D. Mandry a, b, c, D. Leval g, M. Kraemer g
a Imagerie adaptative, diagnostique et interventionnelle, Nancy université, 54511 Vandoeuvre-Lès-Nancy, France 
b Inserm U947, IADI U947, CHU Nancy Brabois, Tour Drouet-4, rue du Morvan, 54511 Vandoeuvre-Lès-Nancy, France 
c Radiologie et medecine nucléaire, CHU de Nancy, 54511 Vandoeuvre-Lès-Nancy, France 
d Inserm, CIC-IT 801, 54511 Nancy, France 
e InESS, université de Strasbourg, 67037 Strasbourg cedex, France 
f UMR 7163, CNRS, 67037 Strasbourg cedex, France 
g Schiller médical, 67160 Wissembourg, France 

Auteur correspondant.

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

pages 6
Iconographies 4
Vidéos 0
Autres 0

Résumé

L’imagerie par résonance magnétique (IRM) cardiaque devient un outil essentiel pour le diagnostic des pathologies cardiaques. Cette technique est très dépendante de la qualité de l’électrocardiogramme (ECG) et de la synchronisation des séquences d’acquisition IRM. Notre objectif a été de développer un capteur ECG « intelligent » totalement compatible avec l’environnement électromagnétique de l’IRM afin de : (1) permettre une synchronisation fiable des séquences IRM en incluant le traitement des troubles du rythme et les patients ayant des ECG de faible amplitude ; et (2) permettre un monitorage continu des patients pour tous types de séquences IRM. Pour atteindre ces objectifs, des capteurs ECG et de mesure de champ magnétique en 3D ont été développés en technologie microélectronique, auxquels nous avons associé un capteur pour la mesure des mouvements respiratoires. Ces derniers représentent correctement la respiration du patient permettant un meilleur contrôle de l’acquisition et de la reconstruction en IRM. Les circuits conçus sont totalement amagnétiques et peuvent donc être placés directement dans le champ de vue de l’IRM. La création d’une base de données ECG a été nécessaire pour mettre en œuvre et évaluer de nouvelles méthodes de traitement de signal comme des techniques de séparation de sources, de décomposition en ondelettes et des approches bayésiennes. Grâce à ces méthodes, il a été possible de supprimer la majorité des artéfacts sur l’ECG pour un meilleur diagnostic et une très bonne synchronisation des séquences. En revanche, les méthodes de classification des troubles du rythme en IRM n’ont pas encore été complètement développées.

Le texte complet de cet article est disponible en PDF.

Abstract

Cardiac magnetic resonance imaging (MRI) has become an essential tool to diagnose several cardiac diseases. This technique is highly dependant on the accuracy of electrocardiogram (ECG) signals and on the sequence synchronization algorithm. Our goal was the development of a smart MR ECG, totally compatible with the MR environment for: (1) robust synchronization of MR sequences, including cases of pathologic ECG, rhythmic disorders, low amplitude ECG for patients who are excluded nowadays and (2) safe and continuous patient monitoring working for all MR examinations with detection of emergency situations. To achieve these objectives, an ECG amplifier and a 3D magnetic field sensor have been developed in microelectronic technology (chip). We have associated motion sensors to acquire respiration displacement during an MRI exam. These sensors give a correct representation of the patient breathing allowing advanced image reconstruction methods. The designed microelectronic circuits are totally non magnetic, and can be placed in the imaging field of view. A dedicated ECG database was created to evaluate and validate new signal processing techniques as, independent component analysis, wavelet decomposition and Bayesian approaches. These methods allow to remove almost all artifacts on the ECG signal to achieve a better diagnosis and smart sequence synchronization. Classification of rhythmic disorder was not totally achieved.

Le texte complet de cet article est disponible en PDF.

Mots clés : Capteurs de mouvements, ECG, IRM, Microélectronique, Microsystème, Monitorage des patients, Synchronisation

Keywords : ECG, Microelectronics, Microsystem, Motion sensor, MRI, Patient monitoring, Synchronization


Plan


 Cet article a été rédigé en vue de la publication dans le numéro spéciale ANR TECSAN « Technologies pour la santé et l’autonomie », volume 32/2 publié en avril 2011.


© 2011  Publié par Elsevier Masson SAS.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 32 - N° 3

P. 179-184 - juin 2011 Retour au numéro
Article précédent Article précédent
  • Mesure de la résistance à l’insuline avec un traceur radioactif du transport du glucose : [123I]-6-Déoxy-6-Iodo-D-Glucose
  • S. Grauzam, M. Henri, L. Slimani, S. Vidal, P. Perret, D. Fagret, F. Boucher, C. Ghezzi
| Article suivant Article suivant
  • Vectorisation et délivrance ciblée de médicaments ou gènes inductibles par des nanoparticules sensibles à l’hyperthermie sous contrôle de l’IRM - NanoBioImaging
  • K.-G. Petry, P. Voisin, A. Brisson, S. Lecommandoux, É. Duguet, J.-M. Franconi, J.-M. Idée, C. Boiziau

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’achat d’article à l’unité est indisponible à l’heure actuelle.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.