Preventing central venous catheter-associated bloodstream infections: Development of an antiseptic barrier cap for needleless connectors - 15/08/11

Doi : 10.1016/j.ajic.2008.10.006 
Steve Z. Menyhay, RN, BSN a, ⁎ , Dennis G. Maki, MD b
a Menyhay Healthcare Systems LLC, Fresno, CA 
b Section of Infectious Diseases, Department of Medicine, University of Wisconsin Hospital and Clinics, Madison, WI 

Address correspondence to Steve Z. Menyhay, RN, BSN, Menyhay Healthcare Systems LLC, Fresno, CA.

Abstract

Background

Reports of outbreaks associated with the use of needle-free valve connectors suggest 2 common risk factors: (1) poor adherence to disinfection practices before use and (2) a design that allows contamination when not in use. Swabbing a membranous septum with 70% isopropyl alcohol may not eliminate septal surface contamination. Frequent access through and handling of needle-free connectors also puts patients at increased risk of central venous catheter-related bloodstream infections (CR-BSIs). A novel antiseptic barrier cap has been designed to maintain health care worker safety and eliminate the vulnerabilities of existing systems to contamination and CR-BSIs for patients at risk.

Methods

A prospective in vitro study compared the effectiveness of standard disinfection of needleless luer-activated valve connectors with 70% isopropyl alcohol and the effectiveness of an antiseptic barrier cap that, when threaded onto a luer-activated connector, rapidly sterilizes a heavily contaminated surface. Standard disinfection was done by 3- to 5-second swabbing using a sterile commercial pledget of 70% isopropyl alcohol. The antiseptic barrier cap comprises an outer cap with internal female threads and a spike inside the closed end, a capsule containing 0.25 mL of 2% chlorhexidine gluconate in 70% isopropyl alcohol, and a sponge between the septum and the capsule. When the cap is threaded onto a luer-adaptable needleless connector, the spike ruptures the capsule, saturating the sponge with the antiseptic.

Results

All 15 (100%) of the precontaminated positive control connectors not disinfected before entry showed transmission of Enterococcus faecalis across the membranous septum (4500-28,000 colony-forming units), and 20 (67%) of 30 connectors disinfected with 70% alcohol showed transmission (442-25,000 colony-forming units). Of 60 needle-free connectors disinfected with the antiseptic barrier cap, 1 (1.6%) showed transmission (P < .001).

Conclusion

An antiseptic barrier cap was highly effective in sterilizing the septum of a needle-free valve connector and preventing entry of any microorganisms, even with heavy contamination of the septum. This new technology should now be evaluated in a clinical trial with CR-BSI as the primary outcome measure.

Le texte complet de cet article est disponible en PDF.

Plan


 Disclosures: S. Z. Menyhay received an honorarium for participating in the symposium and writing this article. Mr. Menyhay is the inventor of the antiseptic barrier cap (Saralex-cl). Menyhay Health care System reports no conflicts of interest.


© 2008  Association for Professionals in Infection Control and Epidemiology, Inc.. Publié par Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 36 - N° 10

P. S174.e1-S174.e5 - décembre 2008 Retour au numéro
Article précédent Article précédent
  • Timing of fluconazole therapy and mortality in patients with Candidemia
  • Kevin W. Garey
| Article suivant Article suivant
  • Clinical use of disinfectable needle-free connectors
  • Juan C. Yébenes, Mateu Serra-Prat

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Elsevier s'engage à rendre ses eBooks accessibles et à se conformer aux lois applicables. Compte tenu de notre vaste bibliothèque de titres, il existe des cas où rendre un livre électronique entièrement accessible présente des défis uniques et l'inclusion de fonctionnalités complètes pourrait transformer sa nature au point de ne plus servir son objectif principal ou d'entraîner un fardeau disproportionné pour l'éditeur. Par conséquent, l'accessibilité de cet eBook peut être limitée. Voir plus

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2026 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.