S'abonner

Intravascular Ultrasound Parameters Associated With Stent Thrombosis After Drug-Eluting Stent Deployment - 16/08/11

Doi : 10.1016/j.amjcard.2007.03.072 
Teruo Okabe, MD a, Gary S. Mintz, MD b, Ashesh N. Buch, MD a, Probal Roy, MD a, Young Joon Hong, MD a, Kimberly A. Smith, BS a, Rebecca Torguson, BS a, Natalie Gevorkian, MD a, Zhenyi Xue, MS a, Lowell F. Satler, MD a, Kenneth M. Kent, MD a, Augusto D. Pichard, MD a, Neil J. Weissman, MD a, Ron Waksman, MD a,
a Cardiovascular Research Institute, Washington Hospital Center, Washington, DC 
b Cardiovascular Research Foundation, New York, New York. 

Corresponding author: Tel.: 202-877-2812; fax: 202-877-2715.

Résumé

Drug-eluting stent (DES) thrombosis (ST) can be devastating. The study aim was to evaluate intravascular ultrasound (IVUS) predictors for DES thrombosis by comparing IVUS studies after implantation in 13 patients with 14 DES thrombosis lesions with a group of controls (30 lesions in 27 patients) matched for history of chronic renal failure and type of DES. Five patients (38%) discontinued dual antiplatelet therapy at the time of ST. There were 3 in-stent restenosis lesions (21%) treated using DESs in the ST group compared with 0 in the control group (p <0.05). Compared with the control group, IVUS studies in the ST group showed a smaller minimum stent area (4.6 ± 1.1 vs 5.6 ± 1.7 mm2, p = 0.0489). In the ST group, 11 of 14 stents had a minimum stent area ≤5.0 mm2 compared with 12 of 30 in the control group (p = 0.0392). Minimum stent area in patients who stopped clopidogrel therapy and developed ST (5.30 ± 1.15 mm2) tended to be larger compared with that in patients who developed ST while using clopidogrel (4.24 ± 0.96 mm2, p = 0.091). Within the 5-mm-long proximal and distal reference segments analyzed, the ST group had larger proximal reference maximum plaque burdens and smaller minimum lumen areas, along with a tendency toward similar findings in the distal reference segments. In conclusion, IVUS findings at the time of DES implantation in patients who subsequently developed ST showed a smaller minimum stent area (especially in patients who developed ST while using clopidogrel) and more residual disease at the stent edges.

Le texte complet de cet article est disponible en PDF.

Plan


© 2007  Elsevier Inc. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 100 - N° 4

P. 615-620 - août 2007 Retour au numéro
Article précédent Article précédent
  • Effects of Triple Antiplatelet Therapy (Aspirin, Clopidogrel, and Cilostazol) on Platelet Aggregation and P-Selectin Expression in Patients Undergoing Coronary Artery Stent Implantation
  • Bong-Ki Lee, Seung-Whan Lee, Seong-Wook Park, Se-Whan Lee, Duk-Woo Park, Young-Hak Kim, Cheol Whan Lee, Myeong-Ki Hong, Jae-Joong Kim, Seongsoo Jang, Hyun-Sook Chi, Seung-Jung Park
| Article suivant Article suivant
  • Meta-Analysis of Angiographic Versus Intravascular Ultrasound Parameters of Drug-Eluting Stent Efficacy (from TAXUS IV, V, and VI)
  • Esteban Escolar, Gary S. Mintz, Jeffrey Popma, Aleksandra Michalek, Sang Wook Kim, Lazar Mandinov, Joerg Koglin, Gregg Stone, Stephen G. Ellis, Eberhard Grube, Keith D. Dawkins, Neil J. Weissman

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Elsevier s'engage à rendre ses eBooks accessibles et à se conformer aux lois applicables. Compte tenu de notre vaste bibliothèque de titres, il existe des cas où rendre un livre électronique entièrement accessible présente des défis uniques et l'inclusion de fonctionnalités complètes pourrait transformer sa nature au point de ne plus servir son objectif principal ou d'entraîner un fardeau disproportionné pour l'éditeur. Par conséquent, l'accessibilité de cet eBook peut être limitée. Voir plus

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2026 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.