EUS-guided radiofrequency ablation with a prototype electrode array system in an animal model (with video) - 23/08/11

Doi : 10.1016/j.gie.2009.03.008 
Shyam Varadarajulu, MD , Nirag C. Jhala, MD, Ernesto R. Drelichman, MD
Current affiliations: Division of Gastroenterology-Hepatology (S.V.), Department of Anatomic Pathology (N.C.J.), and Division of Gastrointestinal Surgery (E.R.D.), University of Alabama at Birmingham School of Medicine, Birmingham, Alabama 

Reprint requests: Shyam Varadarajulu, MD, University of Alabama at Birmingham Medical Center, 410 LHRB, 1530 3rd Avenue South, Birmingham, AL 35294.

Birmingham, Alabama, USA

Abstract

Background

Although previously reported in an animal model, the development of EUS-guided radiofrequency ablation (EUS-RFA) has been impeded because of a lack of a retractable needle electrode array that could safely and effectively ablate large areas.

Objective

To evaluate the feasibility and safety of performing EUS-RFA with a 19-gauge FNA needle fitted with an umbrella-shaped retractable needle electrode array.

Design and Setting

Endoscopic experimental study in a porcine survival model at a tertiary referral center animal laboratory.

Main Outcome Measures

Evaluate the safety and efficacy of the retractable needle electrode array for performing EUS-RFA.

Interventions

A 19-gauge EUS-FNA needle was modified and fitted with a retractable echogenic umbrella-shaped monopolar electrode array at its tip. The FNA needle was connected to a 200-W generator that has an impedance-based feedback system. EUS-RFA of the liver was attempted on 5 Yorkshire pigs. Although 1 pig was euthanized immediately after RFA to assess for immediate complications and pathological examination, the 4 others were kept alive for 7 days.

Results

At EUS, the needle electrode was well visualized and could be deployed in the liver without technical difficulty. During ablation, a round hyperechoic focus gradually surrounded the electrode tip. Tissue ablation was attained within 7 minutes, and the electrode array could be easily withdrawn into the needle assembly. The vital signs of all pigs remained stable throughout the procedure and until they were euthanized. Histopathology in all pigs revealed a discrete, well-demarcated spherical focus of complete coagulation necrosis measuring 2.6 cm in diameter and without damage to the surrounding liver parenchyma or vasculature.

Conclusions

In this experimental study, EUS-RFA of the liver was performed safely by using the retractable umbrella-shaped electrode array with effective coagulation necrosis of large areas.

Le texte complet de cet article est disponible en PDF.

Abbreviations : RF, RFA


Plan


 DISCLOSURE: The following authors disclosed financial relationships relevant to this publication: S. Varadarajulu: Speaker and consultant for Boston Scientific. E. R. Drelichman: Speaker and consultant for Boston Scientific. The other author disclosed no financial relationship relevant to this publication.
 If you want to chat with an author of this article, you may contact him at svaradarajulu@yahoo.com.


© 2009  American Society for Gastrointestinal Endoscopy. Publié par Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 70 - N° 2

P. 372-376 - août 2009 Retour au numéro
Article précédent Article précédent
  • Colonoscopic tips and tricks—advice from 3 master endoscopists
  • Tyler M. Berzin
| Article suivant Article suivant
  • Successful diagnostic and therapeutic intrauterine fetal interventions by natural orifice transluminal endoscopic surgery (with videos)
  • Samuel A. Giday, Jonathan M. Buscaglia, Janyne Althaus, Gianfranco Donatelli, Devi M. Krishnamurty, Xavier Dray, Dawn Ruben, David Liang, Ronald Wroblewski, Priscilla Magno, Eun Ji Shin, Anthony N. Kalloo

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Elsevier s'engage à rendre ses eBooks accessibles et à se conformer aux lois applicables. Compte tenu de notre vaste bibliothèque de titres, il existe des cas où rendre un livre électronique entièrement accessible présente des défis uniques et l'inclusion de fonctionnalités complètes pourrait transformer sa nature au point de ne plus servir son objectif principal ou d'entraîner un fardeau disproportionné pour l'éditeur. Par conséquent, l'accessibilité de cet eBook peut être limitée. Voir plus

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2026 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.