Biophysical characteristics of radiofrequency lesion formation in vivo: Dynamics of catheter tip–tissue contact evaluated by intracardiac echocardiography - 10/09/11

Doi : 10.1016/S0002-8703(97)70242-4 
Jonathan M. Kalman, MBBS, PhD, Adam P. Fitzpatrick, MD, Jeffrey E. Olgin, MD, Michael C. Chin, BS, Randall J. Lee, MD, PhD, Melvin M. Scheinman, MD, Michael D. Lesh, MD

From the Department of Medicine and the Cardiovascular Research Institute, University of California, San Francisco.

San Francisco, Calif. 

Abstract

During clinical radiofrequency catheter ablation a wide range of delivered power may be necessary to achieve success despite an apparently stable catheter position on fluoroscopy. The purpose of this study was to use intracardiac echocardiography to characterize the relation between catheter tip–tissue contact and the efficiency of heating during applications of radiofrequency energy in vivo and to determine whether intracardiac echocardiography could be used prospectively to improve tissue contact. A closed-loop temperature feedback control system was used during radiofrequency applications at five anatomic regions in the right atrium of 15 anesthetized dogs to ensure achievement of a predetermined temperature (70° C) at the catheter tip thermistor by automatic adjustment of delivered power (maximum 100 W). The efficiency-of-heating index was defined as the ratio of steady-state temperature (degrees Celsius) to power (watts). Two-dimensional intracardiac echocardiography was used to evaluate movement of the catheter tip relative to the endocardium. Perpendicular contact was scored as good, average, or poor and lateral catheter sliding as <2, 2 to 5, or >5 mm. Two groups of animals were included: group 1, in which tissue contact was guided by fluoroscopic and electrographic criteria for stability of contact, with intracardiac echocardiography used simply to observe the application; and group 2, in which tissue contact was guided by intracardiac echocardiography. Of 66 applications, 18 (27.3%) had poor perpendicular contact on echocardiography, and 12 (18.2%) demonstrated lateral sliding of >5 mm even though they had been considered to have good tissue contact by fluoroscopic and electrographic criteria. Perpendicular catheter contact and anatomic location were shown to be independently related to the efficiency-of-heating index. Applications with good perpendicular contact had a significantly higher efficiency-of-heating index and a significantly greater lesion size than those with average or poor contact. The percentage of applications having good perpendicular tissue contact and the lesion size were significantly greater when tissue contact was guided by intracardiac echocardiography compared with fluoroscopic and electrographic guidance. This study demonstrates that variations in catheter tip–tissue contact account for differences in the efficiency of tissue heating, independently of the anatomic site of the application. Poor tissue contact was observed by intracardiac echocardiography and confirmed by indexes of tissue heating in approximately one third of radiofrequency applications despite a fluoroscopic appearance and electrographic morphologic appearance suggestive of good tissue contact. There was a significant correlation between echocardiographic evaluation of tissue contact, parameters of tissue heating (efficiency-of-heating index), and lesion size. In addition, intracardiac echocardiography could be used prospectively to improve the percentage of good contact applications and increase the lesion size. (Am Heart J 1997;133:8-18.)

Le texte complet de cet article est disponible en PDF.

Plan


 Dr. Kalman receives funding as the Ralph Reader Overseas Research Fellow of the National Heart Foundation of Australia.
 Reprint requests: Michael D. Lesh, MD, Room MU 428, Box 1354, 500 Parnassus Ave., University of California, San Francisco, CA 94143-1354. E-mail: 7esh@p4.ucsf.edu
 4/1/75871


© 1997  Mosby, Inc. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 133 - N° 1

P. 8-18 - janvier 1997 Retour au numéro
Article précédent Article précédent
  • Are soft echoes really soft? Intravascular ultrasound assessment of mechanical properties in human atherosclerotic tissue
  • Takafumi Hiro, Cyril Y. Leung, Sarah De Guzman, Vincent J. Caiozzo, Ali R. Farvid, Houshang Karimi, Richard H. Helfant, Jonathan M. Tobis
| Article suivant Article suivant
  • Requirement for accurate measurement of regurgitant stroke volume by the combined continuous-wave Doppler and color Doppler flow convergence method
  • Shiro Nozaki, Robin Shandas, Anthony N. DeMaria, From the Division of Cardiovascular Medicine, University of California at San Diego Medical Center.

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Elsevier s'engage à rendre ses eBooks accessibles et à se conformer aux lois applicables. Compte tenu de notre vaste bibliothèque de titres, il existe des cas où rendre un livre électronique entièrement accessible présente des défis uniques et l'inclusion de fonctionnalités complètes pourrait transformer sa nature au point de ne plus servir son objectif principal ou d'entraîner un fardeau disproportionné pour l'éditeur. Par conséquent, l'accessibilité de cet eBook peut être limitée. Voir plus

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2026 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.