Development and Validation of a Noninvasive Method to Determine Arterial Pressure and Vascular Compliance - 10/09/11

Doi : 10.1016/S0002-9149(97)00353-6 
Todd J. Brinton, BS a, Bruno Cotter, MD a, Mala T. Kailasam, MBBS a, David L. Brown, MD a, Shiu-Shin Chio, PhD a, Daniel T. O’Connor, MD a, Anthony N. DeMaria, MD a,
a Department of Medicine, University of California, and Pulse Metric, Incorporated, San Diego, CaliforniaUSA 

*Anthony N. DeMaria, MD, Cardiology Division, UCSD Medical Center, 200 West Arbor Street, San Diego, California 92103-8411.

Abstract

The ability not only to record automated systolic and diastolic pressure, but also to derive measurements of the rate of pressure change during the cardiac cycle, would have great potential clinical value. A new method has been developed to obtain pressure measurements at 20-ms intervals by oscillometric cuff signal pattern recognition. Derivation of noninvasive pressure measurements is based on a T tube aorta and straight tube brachial artery, and assumes that the systolic phase of the suprasystolic cuff signal and the diastolic phase of the subdiastolic cuff signal most closely approximate systolic and diastolic aortic pressures, respectively. Arterial pressures obtained by this method were compared with simultaneous invasive measurements from the thoracic aorta in 36 patients. Good agreement was observed between noninvasive and invasive methods for systolic (146 ± 4 vs 145 ± 5 mm Hg), diastolic (80 ± 2 vs 77 ± 2 mm Hg), and mean (100 ± 3 vs 100 ± 3 mm Hg) arterial pressures, and correlation coefficients were r = 0.94, 0.91, and 0.95, respectively. To assess the validity of measurements of the rate of pressure change, oscillometric cuff signals from a subgroup of 14 patients were analyzed in detail for the peak positive pressure derivative (dP/dtMax), peak negative pressure derivative (dP/dtMin), and time interval between peak positive and peak negative pressure derivatives [tpp]. Results (mean ± SEM) were:

 dP/dtMax dP/dtMin tpp 
Noninvasive 600 ± 41 −466 ± 36 0.25 ± 0.01 
Invasive 508 ± 37 −377 ± 27 0.25 ± 0.01 
r (linear regression) 0.87 0.68 0.85 

The incorporation of measurements of the rate of pressure change into a physical model of the brachial artery was used to derive vascular compliance. A significant correlation was observed between vascular compliance derived from the oscillometric signal and determinations by either thermodilution or Fick methods and noninvasive pressures (n = 20, r = 0.83, p <0.001). Day-to-day variability for blood pressure and vascular compliance derived by the noninvasive method did not differ by >4%, representing a reproducible measure of vascular structure and function. We conclude that the measurement of absolute pressure and rate of pressure change show good correlation with catheter data and that vascular compliance can be reliably assessed by this new method. The technology should provide a valuable noninvasive tool for the assessment of both cardiac function and vascular properties.

Le texte complet de cet article est disponible en PDF.

Plan


© 1997  Elsevier Science Inc. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 80 - N° 3

P. 323-330 - août 1997 Retour au numéro
Article précédent Article précédent
  • Role of Right Ventricular and Pulmonary Functional Abnormalities in Limiting Exercise Capacity in Adults With Congenital Heart Disease
  • Vera H Rigolin, Jennifer S Li, Michael W Hanson, Martin J Sullivan, Paul A Robiolio, Steven E Hearne, William A Baker, J.Kevin Harrison, Thomas M Bashore
| Article suivant Article suivant
  • Effect of Supplemental Oral l-Arginine on Exercise Capacity in Patients With Stable Angina Pectoris
  • Leszek Ceremu?y?ski, Tomasz Chamiec, Krystyna Herbaczy?ska-Cedro

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Elsevier s'engage à rendre ses eBooks accessibles et à se conformer aux lois applicables. Compte tenu de notre vaste bibliothèque de titres, il existe des cas où rendre un livre électronique entièrement accessible présente des défis uniques et l'inclusion de fonctionnalités complètes pourrait transformer sa nature au point de ne plus servir son objectif principal ou d'entraîner un fardeau disproportionné pour l'éditeur. Par conséquent, l'accessibilité de cet eBook peut être limitée. Voir plus

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2026 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.