S'abonner

Randomized trial comparing the 22-gauge aspiration and 22-gauge biopsy needles for EUS-guided sampling of solid pancreatic mass lesions - 20/07/12

Doi : 10.1016/j.gie.2012.03.1392 
Ji Young Bang, MBBS, MPH 1, Shantel Hebert-Magee, MD 2, Jessica Trevino, MD 1, Jayapal Ramesh, MD 1, Shyam Varadarajulu, MD 1,
1 Division of Gastroenterology-Hepatology, University of Alabama at Birmingham, Birmingham, Alabama, USA 
2 Department of Pathology, University of Alabama at Birmingham, Birmingham, Alabama, USA 

Reprint requests: Shyam Varadarajulu, MD, Director of Endoscopy, Basil I. Hirschowitz Endoscopic Center of Excellence, University of Alabama at Birmingham, JT 664, 619-19th Street S, Birmingham, AL 35294-0007

Résumé

Background

To overcome limitations of cytology, biopsy needles have been developed to procure histologic samples during EUS.

Objective

To compare 22-gauge (G) FNA and 22G biopsy needles (FNB) for EUS-guided sampling of solid pancreatic masses.

Design

Randomized trial.

Setting

Tertiary-care medical center.

Patients

This study involved 56 patients with solid pancreatic masses.

Intervention

Sampling of pancreatic masses by using 22G FNA or 22G FNB devices.

Main Outcome Measurements

Compare the median number of passes required to establish the diagnosis, diagnostic sufficiency, technical performance, complication rates, procurement of the histologic core, and quality of the histologic specimen.

Results

A total of 28 patients were randomized to the FNA group and 28 to the FNB group. There was no significant difference in median number of passes required to establish the diagnosis (1 [interquartile range 1-2.5] vs 1 [interquartile range 1-1]; P = .21), rates of diagnostic sufficiency (100% vs 89.3%; P = .24), technical failure (0 vs 3.6%; P = 1.0), or complications (3.6% for both) between FNA and FNB needles, respectively. Patients in whom diagnosis was established in passes 1, 2, and 3 were 64.3% versus 67.9%, 10.7% versus 17.9%, and 25% versus 3.6%, respectively, for FNA and FNB cohorts. There was no significant difference in procurement of the histologic core (100% vs 83.3%; P = .26) or the presence of diagnostic histologic specimens (66.7% vs 80%; P = .66) between FNA and FNB cohorts, respectively.

Limitations

Only pancreatic masses were evaluated.

Conclusion

Diagnostic sufficiency, technical performance, and safety profiles of FNA and FNB needles are comparable. There was no significant difference in yield or quality of the histologic core between the 2 needle types. (Clinical trial registration number: NCT01394159.)

Le texte complet de cet article est disponible en PDF.

Abbreviations : EUS-FNA, FNB, HBSS, TNB


Plan


 DISCLOSURE: S. Varadarajulu is a consultant for Boston Scientific. No other financial relationships relevant to this publication were disclosed.
 See CME section; p. 396.


© 2012  American Society for Gastrointestinal Endoscopy. Publié par Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 76 - N° 2

P. 321-327 - août 2012 Retour au numéro
Article précédent Article précédent
  • Diagnostic significance of a dilated orifice of the duodenal papilla in intraductal papillary mucinous neoplasm of the pancreas
  • Teppei Aso, Takao Ohtsuka, Noboru Ideno, Hiroshi Kono, Yosuke Nagayoshi, Yasuhisa Mori, Kennoki Ohuchida, Junji Ueda, Shunnichi Takahata, Katsuya Morimatsu, Shinichi Aishima, Hisato Igarashi, Tetsuhide Ito, Kousei Ishigami, Kazuhiro Mizumoto, Masao Tanaka
| Article suivant Article suivant
  • Diagnostic yield of malignancy during EUS-guided FNA of solid lesions with and without a stylet: a prospective, single blind, randomized, controlled trial
  • Sachin Wani, Dayna Early, Julie Kunkel, Ann Leathersich, Christine E. Hovis, Thomas G. Hollander, Cara Kohlmeier, Cynthia Zelenka, Riad Azar, Steven Edmundowicz, Brian Collins, Jingxia Liu, Matthew Hall, Daniel Mullady

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Elsevier s'engage à rendre ses eBooks accessibles et à se conformer aux lois applicables. Compte tenu de notre vaste bibliothèque de titres, il existe des cas où rendre un livre électronique entièrement accessible présente des défis uniques et l'inclusion de fonctionnalités complètes pourrait transformer sa nature au point de ne plus servir son objectif principal ou d'entraîner un fardeau disproportionné pour l'éditeur. Par conséquent, l'accessibilité de cet eBook peut être limitée. Voir plus

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2026 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.