Résumé

Background

Computer-based endoscopy simulators may enable trainees to learn and develop technical skills before performing on patients. Simulators require validation as adequate models of live endoscopy before being used for training or assessment purposes.

Objective

To evaluate content and criterion validity of the CAE EndoscopyVR Simulator colonoscopy and EGD modules as predictors of clinical endoscopic skills.

Design

Prospective, observational, non-randomized, parallel cohort study.

Setting

Single academic center with accredited gastroenterology training program.

Participants

Five novice first-year gastroenterology fellows and 6 expert gastroenterology attending physicians.

Intervention

Participants performed 18 simulated colonoscopies and 6 simulated EGDs. The simulator recorded objective performance parameters. Participants then completed feedback surveys.

Main Outcome Measurements

The 57 objective performance parameters measured by the endoscopy simulator were compared between the two study groups. Novice and expert survey responses were analyzed.

Results

Significant differences between novice and expert performance were detected in only 19 of 57 (33%) performance metrics. Eight of these 19 (42%) were time-related metrics, such as total procedure time, time to anatomic landmarks, and time spent in contact with GI mucosa. Of 49 non-time related measures, the few additional statistically significant differences between novices and experts involved air insufflation, sedation management, endoscope force, and patient comfort. These findings are of uncertain clinical significance. Survey data found multiple aspects of the simulation to be unrealistic compared with human endoscopy.

Limitations

Small sample size.

Conclusion

The CAE EndoscopyVR Simulator displays poor content and criterion validity and is thereby incapable of predicting skill during in vivo endoscopy.

Le texte complet de cet article est disponible en PDF.

Plan


 DISCLOSURE: M.L. Kochman received grant support from the Greenberg-Gringlas Fund. No other financial relationships relevant to this publication were disclosed.


© 2012  American Society for Gastrointestinal Endoscopy. Publié par Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 76 - N° 5

P. 993-1002 - novembre 2012 Retour au numéro
Article précédent Article précédent
  • Diagnostic yield of small-bowel capsule endoscopy in patients with iron-deficiency anemia: a systematic review
  • Anastasios Koulaouzidis, Emanuele Rondonotti, Andry Giannakou, John N. Plevris
| Article suivant Article suivant
  • Pathological reassessment of hyperplastic colon polyps in a city-wide pathology practice: implications for polyp surveillance recommendations
  • Harminder Singh, Diane Bay, Stephen Ip, Charles N. Bernstein, Zoann Nugent, Rodica Gheorghe, Robert Wightman

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Elsevier s'engage à rendre ses eBooks accessibles et à se conformer aux lois applicables. Compte tenu de notre vaste bibliothèque de titres, il existe des cas où rendre un livre électronique entièrement accessible présente des défis uniques et l'inclusion de fonctionnalités complètes pourrait transformer sa nature au point de ne plus servir son objectif principal ou d'entraîner un fardeau disproportionné pour l'éditeur. Par conséquent, l'accessibilité de cet eBook peut être limitée. Voir plus

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2026 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.