Optimum cystic duct closure: a comparative study using metallic clips, ENSEAL, and ENDOLOOP in swine - 14/04/13

Doi : 10.1016/j.amjsurg.2013.01.011 
Derek Mcvay, D.O., Daniel Nelson, D.O., Christopher R. Porta, M.D., Kelly Blair, M.D., Matthew Martin, M.D.
Department of Surgery, Madigan Healthcare System, Madigan Army Medical Center, 9040-A Fitzsimmons Avenue, Tacoma, WA 98431, USA 

Corresponding author. Tel.: +1-253-968-2361; fax: +253/968-5900.

Abstract

Background

Metal clips are commonly used to secure the cystic duct during cholecystectomy, although use of an ENDOLOOP (Ethicon Endo-Surgery, Blue Ash, OH) is often touted as a more secure closure when postoperative endoscopic retrograde cholangiopancreatography (ERCP) is anticipated. The objective of this study was to test the strength of 3 different cystic duct closure methods in a model simulating postoperative biliary insufflation.

Methods

The extrahepatic biliary system, including common bile duct, gallbladder, and cystic duct, was harvested en bloc from 22 swine postmortem. A cholecystectomy was performed and the cystic duct was secured using 1 of 3 randomly assigned methods: metallic clips (Ethicon Endo-Surgery), an ENDOLOOP (Ethicon Endo-Surgery), or an ENSEAL tissue sealing device (Ethicon Endo-Surgery). The common bile duct was cannulated with a pressure-monitoring system and insufflated with air. The burst pressures, location of rupture, and size of the common bile duct and cystic duct were recorded and compared.

Results

There were 7 pigs each in the ENDOLOOP and ENSEAL groups and 8 in the metallic clip group, with no statistical significance between cystic and common bile duct size. Mean burst pressure was 432 mm Hg for metallic clips, 371 mm Hg for the ENDOLOOP, and 238 mm Hg for the ENSEAL device (P = .02). Post hoc analysis revealed clips to be statistically superior when compared with the ENSEAL (P= .01). There was no statistical difference between the ENDOLOOP and metal clips or between the ENDOLOOP and the ENSEAL.

Conclusions

All 3 closure methods successfully secured the cystic duct, with mean burst pressures exceeding 195 mm Hg. Metallic clips demonstrated the highest burst pressures and no cystic duct stump leaks. This study challenges the traditional dogma of additionally securing the cystic duct with an ENDOLOOP when postoperative biliary instrumentation is expected and also suggests that an adequately secure closure may be obtained with thermal sealing devices.

Le texte complet de cet article est disponible en PDF.

Keywords : Cholecystectomy, Cystic duct, Clips, ENDOLOOP, Energy device, ENSEAL


Plan


 The results and opinions expressed in this article are those of the authors, and do not reflect the opinions or official policy of the United States Army or the Department of Defense.
 ENSEAL devices were provided by Ethicon Endo-Surgery, Blue Ash, OH.
 The authors declare no conflicts of interest.


© 2013  Publié par Elsevier Masson SAS.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 205 - N° 5

P. 547-551 - mai 2013 Retour au numéro
Article précédent Article précédent
  • Perceptions of conflict of interest: surgeons, internists, and learners compared
  • Christopher J. de Gara, Kim C. Rennick, John Hanson
| Article suivant Article suivant
  • Use of volunteer student abstractors for a retrospective cohort analysis: a study of inter-rater reliability
  • Yaroslav Gritsiouk, Damian Hegsted, Stuart Gardiner, Lisa Merriman, Kelly Dean Gubler

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Elsevier s'engage à rendre ses eBooks accessibles et à se conformer aux lois applicables. Compte tenu de notre vaste bibliothèque de titres, il existe des cas où rendre un livre électronique entièrement accessible présente des défis uniques et l'inclusion de fonctionnalités complètes pourrait transformer sa nature au point de ne plus servir son objectif principal ou d'entraîner un fardeau disproportionné pour l'éditeur. Par conséquent, l'accessibilité de cet eBook peut être limitée. Voir plus

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2026 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.