Bacteraemia in Ventricular Assist Devices: A Common Complication that Need Not Affect Clinical Outcomes - 25/02/14

Doi : 10.1016/j.hlc.2013.09.008 
Franklin L. Rosenfeldt, MD, FRACS a, ⁎ , Lachlan J. Kwa, MBBS a, Pramote Porapakkham, MBBS a, Suraindra Rajadurai, MBBS a, Kylie Jones, CCRN a, Juliana van de Merwe, RN a, Baki Billah, PhD c, Pornwalee Porapakkham, MBBS c, Donald S. Esmore, FRACS a, Dag S. Halvorsen, MBBS b, Victor J. Aguirre, MBBS a, Denis W. Spelman, FRACP b
a Department of Cardiothoracic Surgery, The Alfred Hospital, Melbourne, Australia 
b Department of Microbiology and Infectious Diseases, Alfred Hospital, Melbourne, Australia 
c Department of Epidemiology and Preventive Medicine, Monash University, Melbourne, Australia 

⁎Corresponding author. Tel.: +61 3 9276 3685; fax: +61 3 9276 2317.

Résumé

Background

Ventricular assist device (VAD) implantation has become an effective option for patients with severe heart failure. However, device-related infections remain a significant problem. The aim of this study was to describe the incidence and microbiological aetiology of bacteraemia in patients with VADs, and to assess the impact of bacteraemia on clinical outcomes.

Methods

A retrospective study was conducted of patients having VAD implantation at the Alfred Hospital (Melbourne, Australia) from October 1990 to July 2009. Medical records and microbiology databases were reviewed. Patients who were supported with a VAD for 72h or more were evaluated for demographic data, VAD type, the occurrence of bacteraemia and clinical outcomes.

Results

During the 19-year period, 135 VAD patients (89 Thoratec PVAD, 10 Novacor, and 36 Ventrassist) supported for a total duration of 17,304 (median 74) support days were included. Sixty-one patients (45%) developed VAD-associated bacteraemia, an incidence of 5.6 episodes per 1000 support days. The incidence of bacteraemia per 1000 days of support was similar for the three devices used: Thoratec PVAD, Novacor and Ventrassist VADs (7.8±0.8, 5.2±1.5 and 3.4±0.5, respectively, p=0.74). Staphylococcus aureus was the most common pathogen (25%). The rates of death on device, survival to transplant, recovery with explant and outcomes after transplantation, including 30-day mortality, median survival time and incidence of cerebrovascular accidents were not significantly impacted upon by bacteraemia.

Conclusions

Bacteraemia is common in VAD patients. However, the incidence of VAD-associated bacteraemia is independent of device type and with aggressive antimicrobial therapy; clinical outcomes need not be affected by the bacteraemia.

Le texte complet de cet article est disponible en PDF.

Keywords : Ventricular assist device, Bacteraemia, Bridge to transplantation, Destination therapy, Transplantation


Plan


© 2013  Australian and New Zealand Society of Cardiac and Thoracic Surgeons (ANZSCTS) and the Cardiac Society of Australia and New Zealand (CSANZ). Publié par Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 23 - N° 3

P. 234-241 - mars 2014 Retour au numéro
Article précédent Article précédent
  • Hepatic and Renal Function with Successful Long-term Support on a Continuous Flow Left Ventricular Assist Device
  • Salil V. Deo, Vikas Sharma, Salah E. Altarabsheh, Tal Hasin, John Dillon, Ishan K. Shah, Lucian A. Durham, John M. Stulak, Richard C. Daly, Lyle D. Joyce, Soon J. Park
| Article suivant Article suivant
  • Surgical Aortic Valve Replacement in Very Elderly Patients Aged 80 Years and Over: Evaluation of Early Clinical Outcomes
  • Edwin Ho, Manu N. Mathur, Peter W. Brady, David Marshman, Russell J. Brereton, Donald E. Ross, Ravinay Bhindi, Peter S. Hansen

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Elsevier s'engage à rendre ses eBooks accessibles et à se conformer aux lois applicables. Compte tenu de notre vaste bibliothèque de titres, il existe des cas où rendre un livre électronique entièrement accessible présente des défis uniques et l'inclusion de fonctionnalités complètes pourrait transformer sa nature au point de ne plus servir son objectif principal ou d'entraîner un fardeau disproportionné pour l'éditeur. Par conséquent, l'accessibilité de cet eBook peut être limitée. Voir plus

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2026 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.