Acute echocardiographic optimization of multiple stimulation configurations of cardiac resynchronization therapy through quadripolar left ventricular pacing: A tailored approach - 22/03/14

Doi : 10.1016/j.ahj.2013.12.028 
Leonardo Calò, MD, FESC , Annamaria Martino, MD, Ermenegildo de Ruvo, MD, Monia Minati, MD, Simona Fratini, MD, PhD, Marco Rebecchi, MD, Chiara Lanzillo, MD, PhD, Alessandro Fagagnini, MD, Alessio Borrelli, MD, Lucia De Luca, MD, PhD, Luigi Sciarra, MD
 Division of Cardiology, Policlinico Casilino, ASL Rome B, Rome, Italy 

Reprint requests: Leonardo Calò, MD, FESC, Via Buonarroti 16, 00047 Marino, Rome, Italy.

Résumé

Background

Cardiac resynchronization therapy (CRT) is ineffective in approximately 30% of recipients, in part due to sub-optimal left ventricular (LV) pacing location. The Quartet LV lead, with 2 additional electrodes proximal to conventional bipolar lead electrodes, enables 10 different pacing configurations at four independent LV locations. In a CRT patient cohort, we sought to evaluate the spectrum of echocardiographic and electrocardiographic response over these 10 configurations, to select the optimal one in each patient. Moreover, we sought to evaluate the 6-months clinical and echocardiographic response to a “tailored approach” in which the optimal LV pacing configuration for CRT was determined by echocardiographic measures, QRSd and pacing capture thresholds.

Methods

Twenty-two consecutive CRT indicated patients were implanted with a quadripolar CRT system (St. Jude Medical). Optimal LV pacing configuration was determined by echocardiographic measures, including velocity time integral (VTI), myocardial performance index (MPI) and mitral regurgitation (MR), and an electrocardiographic measure (QRS duration) during pacing from each of the configurations at pre-discharge. The optimal LV pacing vector was chosen for every patient. Clinical and echocardiographic assessment was repeated after 6 months.

Results

Various configurations provided different VTI, MPI, MR and QRSd values. Conventional bipolar vectors (ie, D1-M2, D1-RVc, M2-RVc) were rarely associated with the best echocardiographic improvements and provided significantly worse VTI, MR, MPI, and QRSd values than the best configuration for every patient (P = .005, P = .05 and P = .03 for VTI; P = .01, P = .005 and P = .001 for MPI; P = .003, P = .01 and P = .005 for MR, P > .5, P = .01 and P = .05 for QRSd) Conversely, “unconventional” proximal configurations (ie, making use of P4 and M3 electrodes) were generally characterized by higher acute VTI, MR and MPI improvements. CRT devices were reprogrammed with an “unconventional” LV pacing configuration in 50% of patients. A significant improvement in New York Heart Association class (81%), LV ejection fraction (76%), end-diastolic and end-systolic volumes was observed after 6 months (P = .02, P < .001, P = .02 and P = .003, respectively).

Conclusions

In this study, conventional bipolar vectors of quadripolar-CRT were rarely associated with the best echocardiographic improvements. Quadripolar CRT utilizing optimal LV pacing configuration was associated with a significant improvement in New York Heart Association class and LV ejection fraction after 6 months.

Le texte complet de cet article est disponible en PDF.

Plan


© 2014  Mosby, Inc. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 167 - N° 4

P. 546-554 - avril 2014 Retour au numéro
Article précédent Article précédent
  • Scar tissue–guided left ventricular lead placement for cardiac resynchronization therapy in patients with ischemic cardiomyopathy: An acute pressure-volume loop study
  • Gerben Jan de Roest, LiNa Wu, Carel C. de Cock, Matthijs L. Hendriks, Peter Paul H.M. Delnoy, Albert C. van Rossum, Cornelis P. Allaart
| Article suivant Article suivant
  • Ventricular geometric characteristics and functional benefit of mild right ventricular outflow tract obstruction in patients with significant pulmonary regurgitation after repair of tetralogy of Fallot
  • Chun-An Chen, Ssu-Yuan Chen, Jou-Kou Wang, Wen-Yih Isaac Tseng, Hsin-Hui Chiu, Chung-I Chang, Ing-Sh Chiu, Yih-Sharng Chen, Ming-Chun Yang, Chun-Wei Lu, Ming-Tai Lin, Mei-Hwan Wu

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Elsevier s'engage à rendre ses eBooks accessibles et à se conformer aux lois applicables. Compte tenu de notre vaste bibliothèque de titres, il existe des cas où rendre un livre électronique entièrement accessible présente des défis uniques et l'inclusion de fonctionnalités complètes pourrait transformer sa nature au point de ne plus servir son objectif principal ou d'entraîner un fardeau disproportionné pour l'éditeur. Par conséquent, l'accessibilité de cet eBook peut être limitée. Voir plus

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2026 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.