S'abonner

Initial Experience with a Novel Real-Time Three-Dimensional Intracardiac Ultrasound System to Guide Percutaneous Cardiac Structural Interventions: A Phase 1 Feasibility Study of Volume Intracardiac Echocardiography in the Assessment of Patients with Structural Heart Disease Undergoing Percutaneous Transcatheter Therapy - 28/08/14

Doi : 10.1016/j.echo.2014.04.022 
Frank E. Silvestry, MD, FACC, FASE , Mitul B. Kadakia, MD, Judi Willhide, RN, Howard C. Herrmann, MD, FACC, FSCAI
 Cardiovascular Division, Department of Medicine, Hospital of the University of Pennsylvania, Philadelphia, Pennsylvania 

Reprint requests: Frank E. Silvestry, MD, FACC, FASE, University of Pennsylvania School of Medicine, Cardiovascular Division, 9007 E Gates Pavilion, Hospital of the University of Pennsylvania, 3400 Spruce Street, Philadelphia, PA 19104.

Abstract

Background

Intracardiac echocardiographic (ICE) imaging is a modality increasingly used to guide percutaneous cardiac structural interventions. Until recently, ICE imaging has been limited by the presence of only two-dimensional imaging planes and requires considerable catheter manipulation to visualize certain targets. The aim of this study was to assess the feasibility of a new three-dimensional (3D) volumetric ICE system to provide imaging guidance in 15 patients undergoing percutaneous cardiac structural interventions.

Methods

The Siemens AcuNav 3D volumetric ICE catheter was used to guide interventions in 15 patients. Imaging was performed at 6 and 8 MHz without color Doppler flow mapping and at 4 and 6 MHz with color Doppler flow mapping. The images were independently reviewed, and the ability to visualize specific structures was assessed by two independent and expert ICE imagers.

Results

The majority of patients (n = 11 [73%]) were undergoing percutaneous transcatheter closure of patent foramen ovales (n = 3 [20%]) or atrial septal defects (n = 8 [53%]). Three patients (20%) underwent balloon valvuloplasty for mitral stenosis. One patient (7%) underwent a diagnostic study for congenital heart disease. There were no significant differences in image scores between 3D and two-dimensional imaging without color Doppler in clinically important targets. With color Doppler, there were decreased image scores in the 3D images. Three-dimensional images provided improved imaging of devices and catheters and of the relationship between atrial septal defect devices and the aorta.

Conclusion

Three-dimensional volumetric ICE imaging can be successfully used to guide structural heart disease procedures. It has the potential to provide greater anatomic information during interventions. Further improvement in its imaging capabilities is required to improve color Doppler mapping and volume size capabilities.

Le texte complet de cet article est disponible en PDF.

Keywords : Intracardiac echocardiography, Structural heart disease intervention, Mitral valve disease, Atrial septal defect, Patent foramen ovale

Abbreviations : ASD, DICOM, ICE, PFO, TEE, 3D, 2D


Plan


 This study was funded by a grant to the institution from Siemens Medical, Inc (Malvern, PA). Dr Herrmann is a consultant to Siemens Medical, Inc.
 ClinicalTrials.gov Identifier: NCT01669551


© 2014  American Society of Echocardiography. Publié par Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 27 - N° 9

P. 978-983 - septembre 2014 Retour au numéro
Article précédent Article précédent
  • Systolic and Diastolic Function of the Fetal Single Left Ventricle
  • Paul A. Brooks, Nee S. Khoo, Lisa K. Hornberger
| Article suivant Article suivant
  • Intracardiac Echocardiography during Atrial Septal Defect and Patent Foramen Ovale Device Closure in Pediatric and Adolescent Patients
  • Beth A. Medford, Nathaniel W. Taggart, Allison K. Cabalka, Frank Cetta, Guy S. Reeder, Donald J. Hagler, Jonathan N. Johnson

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Elsevier s'engage à rendre ses eBooks accessibles et à se conformer aux lois applicables. Compte tenu de notre vaste bibliothèque de titres, il existe des cas où rendre un livre électronique entièrement accessible présente des défis uniques et l'inclusion de fonctionnalités complètes pourrait transformer sa nature au point de ne plus servir son objectif principal ou d'entraîner un fardeau disproportionné pour l'éditeur. Par conséquent, l'accessibilité de cet eBook peut être limitée. Voir plus

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2026 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.