Morphometric analysis in three-dimensional intracoronary ultrasound: An in vitro and in vivo study performed with a novel system for the contour detection of lumen and plaque - 11/09/11

Doi : 10.1016/S0002-8703(96)90233-1 
Clemens von Birgelen, MD, Carlo Di Mario, MD, PhD, Wenguang Li, MSc, Johan C.H. Schuurbiers, BSc, Cornelis J. Slager, MSc, Pim J. de Feyter, MD, PhD, Jos R.T.C. Roelandt, MD, PhD, Patrick W. Serruys, MD, PhD
Thoraxcenter, Division of Cardiology, University Hospital Rotterdam, Dijkzigt, Erasmus University, Rotterdam, The Netherlands 

Reprint requests: Patrick W. Serruys, MD, PhD, Thoraxcenter, Interventional Cardiology, P.O. Box 1738, University Hospital Rotterdam, Dijkzigt, 3000 DR Rotterdam, The Netherlands.

Abstract

Currently, automated systems for quantitative analysis by intracoronary ultrasound (ICUS) are restricted to the detection of the lumen. The aim of this study was to determine the accuracy and reproducibility of a new semiautomated contour detection method, providing off-line identification of the intimal leading edge and external contour of the vessel in three-dimensional ICUS. The system allows cross-sectional and volumetric quantification of lumen and of plaque. It applies a minimum-cost algorithm and the concept that edge points derived from previously detected longitudinal contours guide and facilitate the contour detection in the cross-sectional images. A tubular phantom with segments of various luminal dimensions was examined in vitro during five catheter pull-backs (1 mm/sec), and subsequently 20 diseased human coronary arteries were studied in vivo with 2.9F 30 MHz mechanical ultrasound catheters (200 images per 20 mm segment). The ICUS measurements of phantom lumen area and volume revealed a high correlation with the true phantom areas and volumes (r = 0.99); relative mean differences were −0.65% to 3.86% for the areas and 0.25% to 1.72% for the volumes of the various segments. Intraobserver and interobserver comparisons showed high correlations (r = 0.95 to 0.98 for area and r = 0.99 for volume) and small mean relative differences (−0.87% to 1.08%), with SD of lumen, plaque, and total vessel measurements not exceeding 7.28%, 10.81%, and 4.44% (area) and 2.66%, 2.81%, and 0.67% (volume), respectively. Thus the proposed analysis system provided accurate measurements of phantom dimensions and can be used to perform highly reproducible area and volume measurements in three-dimensional ICUS in vivo.

Il testo completo di questo articolo è disponibile in PDF.

© 1996  Pubblicato da Elsevier Masson SAS.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 132 - N° 3

P. 516-527 - settembre 1996 Ritorno al numero
Articolo precedente Articolo precedente
  • Differential characterization of human coronary collateral blood flow velocity
  • Christophe Tron, Thomas J. Donohue, Richard G. Bach, Thomas Wolford, Eugene A. Caracciolo, Frank V. Aguirre, Alexander Khoury, Morton J. Kern
| Articolo seguente Articolo seguente
  • Myocardial contrast echocardiography in human beings: Correlation of resting perfusion defects to sestamibi single photon emission computed tomography
  • Mario F. Meza, Sameh Mobarek, Robert Sonnemaker, Stanton Shuler, Stephen R. Ramee, Tyrone J. Collins, Christopher J. White, Dagnovar Aristizabal, Joseph P. Murgo, Jorge Cheirif

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
L'accesso al testo integrale di questo articolo richiede un abbonamento.

Già abbonato a @@106933@@ rivista ?

@@150455@@ Voir plus

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2026 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.