012 - Effect of nebulizer device set on measurement of tidal volume and pressure ventilator during two mode of ventilation - 05/12/08

Doi : RMR-11-2008-25-9-0761-8425-101019-200810877 

E. Mercier,

S. Leguellec,

P. Diot,

L. Vecellio

Voir les affiliations

Bienvenue sur EM-consulte, la référence des professionnels de santé.
Article gratuit.

Connectez-vous pour en bénéficier!

Rationale: Numerous factors affect aerosolized drug delivery during invasive ventilation. However, there is few data concerning measurement of tidal volume and pressures on the ventilator and similarly on the lung during a nebulization.

Aim: To evaluate the effect of aerosol devices in two modes of mechanical ventilation on the measurement of tidal volume and pressures on the adult and infant test lung.

Methods: A Servo 300 and a Galileo ventilator were set for Volume assisted Control (VAC) and Pressure Control (PC) and were connected to a Michigan test lung. Albuterol sulfate inhalation solution was nebulized with 5 aerosol devices: MistyNeb and Sidestream jet nebulizer, Aeroneb Pro vibrating aperture plate nebulizer, Ring nebulizer and AMGH ultrasonic nebulizer. Tidal volumes were set respectively at 450 ml and 180 ml for the adult and infant test lung. Frequency was 15/min in adult and 20/min in infant. Tidal volumes, peak pressure, PEEP were measured on the ventilator and on the test lung before and during nebulization.

Results: Graph 1 shows the pressure curves by test devices and ventilation modes.

Jet nebulizers: During VAC, there was no significant difference on the pressure measurement on the ventilator and the test lung before and during nebulization. The tidal volume measurement increased to 28% in adult and 45% in infant. In PC, tidal volume was not modified and pressure measurements were increased during nebulization.

Ultrasonic nebulizer: During VAC, measurements of pressure and tidal volume were similar with or without nebulizer. During VP, the tidal volume measurement was decreased at 32% in adult and 13% in infant.

Aeroneb Pro and Ring nebulizer: tidal volume and pressure measurements were comparable before and during nebulization.

Conclusion: Aerosol device influenced significantly pressure or volume measurement during mechanical ventilation. These variations of pressure or volume could be dangerous for the patients and incite to choice the optimal ventilation mode for an aerosol device.

Pressure and volumes measurement were not affected by use of Aeroneb Pro and Ring nebulizers during mechanical ventilation and allow a safe clinical practice.




© 2008 Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 25 - N° 9

P. 1159 - novembre 2008 Retour au numéro
Article précédent Article précédent
  • 011 - Caractérisation des cellules basales de l’épithélium bronchique humain adulte
  • T. Jolly, J. Roux, S. Lingee, P. Birembaut, C. Coraux
| Article suivant Article suivant
  • 013 - Association of ex vivo vascular and bronchial dysfunctions in mild COPD
  • M. Oualha, J.-F. Boitiaux, J.-M. Tadie, A. Cazes, M. Riquet, D. Israel-Biet, C. Delclaux

Bienvenue sur EM-consulte, la référence des professionnels de santé.

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.