Effects of autoclaving versus cold chemical (glutaraldehyde) sterilization on load-deflection characteristics of aesthetic coated archwires - 22/05/18

Doi : 10.1016/j.ortho.2018.03.009 
Seyed Mohammad Mousavi a, Elham Hormozi b, , Mina Moradi c, Milad Shamohammadi a, Vahid Rakhshan d
a Department of Orthodontics, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran 
b Department of Orthodontics, Dental Research Center, School of Dentistry, Isfahan University of Medical Sciences, Isfahan, Iran 
c Department of Orthodontics, North Khorasan University of Medical Sciences, Bojnurd, Iran 
d Department of Dental Anatomy, Faculty of Dentistry, Azad University, Tehran, Iran 

Correspondence and reprints: Department of Orthodontics, Dental Research Center, School of Dentistry, Isfahan University of Medical Sciences, Hezar-Jerib Ave., Isfahan, 81746 73461, Iran.Department of Orthodontics, Dental Research Center, School of Dentistry, Isfahan University of Medical Sciences, Hezar-Jerib Ave., Isfahan, 81746 73461, Iran.

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

pages 13
Iconographies 2
Vidéos 0
Autres 0

Summary

Introduction

The effect of any sterilization methods (cold [chemical] or hot) on load-deflection characteristics of aesthetic coated archwires has not yet been investigated. Thus, we assessed it.

Methods

In this experimental in vitro study, 90-coated archwires from 3 brands were purchased. Ten wires from each brand (n=30) underwent cold and 10 underwent hot sterilization, while 10 left unsterilized as negative controls. Load-deflection curves were established for each wire (as five 0.2mm intervals between 1.0 and 1.8mm displacements), using a three-bracket test. After determining the plateau phase that was present in all wires, the averages of plateau phase pertaining to loading and unloading curves and their discrepancies (hysteresis) were calculated for each subgroup (n=10). Average plateau values were compared using two-way ANOVA, Tukey, and independent-samples t-test (α=0.05).

Results

Average loading and unloading values were 906.6±129.7 and 295.9±84.5g respectively (t-test P<0.0001). Comparing loading plateaus, ANOVA indicated significant difference among wire brands (P<0.0001) but not among sterilization types (P=0.4793). Comparing unloading plateaus, ANOVA showed significant differences among wire brands (P<0.0001) and sterilization types (P=0.0008). Tukey showed that only cold sterilization and negative control differed significantly (P<0.001); cold and hot sterilization methods, or control and autoclaving were not significantly different (P>0.05). Comparing hysteresis plateaus, ANOVA indicated difference among wire brands (P<0.0053) but not among sterilization types (P=0.9166).

Conclusions

Cold sterilization might reduce unloading plateau of orthodontic wires, but sterilization in general might not affect loading or hysteresis plateaus. Different brands had different plateaus of loading, unloading and hysteresis.

Le texte complet de cet article est disponible en PDF.

Keywords : Sterilization, Autoclaving, Glutaraldehyde, Load-deflection characteristics


Plan


© 2018  Publié par Elsevier Masson SAS.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 16 - N° 2

P. 281-293 - juin 2018 Retour au numéro
Article précédent Article précédent
  • Orthodontie de camouflage assistée par corticotomie chez un patient adulte de classe III avec déficit transversal sévère
  • Antonio Gracco, Marco Finotti, Giovanni Bruno, Alberto de Stefani
| Article suivant Article suivant
  • Effets des stérilisations à l’autoclave et chimique à froid (glutaraldéhyde) sur les caractéristiques contraintes-déformations des fils orthodontiques esthétiques : étude comparative in vitro
  • Seyed Mohammad Mousavi, Elham Hormozi, Mina Moradi, Milad Shamohammadi, Vahid Rakhshan

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’achat d’article à l’unité est indisponible à l’heure actuelle.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.