RapidRhino herniation causing airway obstruction - 22/06/18
, Alex Paspaliaris, MBBS a, cAbstract |
Epistaxis is commonly managed with RapidRhino in emergency departments. We report a case of RapidRhino herniation during air inflation in a 99-year-old female, leading to significant airway obstruction. Upon re-inflation of the RapidRhino for persistent epistaxis, our patient suddenly developed acute respiratory distress. Throat examination revealed a large clot-like circular mass occluding the oropharynx. Initially mistaken as a massive blood clot, the mass was identified as the distal end of the RapidRhino. Immediate deflation of the RapidRhino results in spontaneous resolution of the respiratory distress. We hypothesise that the RapidRhino has herniated posteriorly into the oropharynx through excessive air inflation, potentially compounded by a product defect. Our case highlights a fatal consequence of RapidRhino application where emergency physicians should be cautious of, especially among inexperienced users.
Le texte complet de cet article est disponible en PDF.Keywords : Airway obstruction, Epistaxis, Tampons, surgical
Plan
| ☆ | The authors have no financial support or relationships concerned in this case report. |
Vol 36 - N° 7
P. 1322.e3-1322.e4 - juillet 2018 Retour au numéroBienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.