Objectives

To describe step-by-step the port placement and the robot docking of the new purpose-built robotic platform for R-LESS. The feasibility of different approaches to the pelvic fossa and the retroperitoneum was reported in cadaver models.

Methods

This was a preclinical study on human cadavers to assess the feasibility of the da Vinci SP1098 surgical system for R-LESS pelvic fossa and retroperitoneal urological surgeries. We used the SP1098 to perform R-LESS prostatectomies and cystoprostatectomies with transperineal and transvesical approaches, and nephrectomies (radical or partial) with retroperitoneal approach. The primary outcome was to report the port placement and docking. The technical feasibility of the procedures was then demonstrated as measured by the need for adjunctive ports or the occurrence of intraoperative complications. Operative times were recorded.

Results

A total of 14 procedures were performed on 12 human cadavers. Namely 4 prostatectomies and 2 cystoprostatectomies with transperineal approach, 3 transvesical prostatectomies, 1 retroperitoneal radical, and 4 retroperitoneal partial nephrectomies. Operative times were in line with those of standard multiport robotic surgery. Neither additional ports nor percutaneous instruments were required. No intraoperative complications occurred. Limitations include the preclinical model, the small sample size, and the lack of a control group.

Conclusion

In this preclinical model, the port placement and robot docking using the SP1098 robotic platform is reproducible and feasible for pelvic fossa and retroperitoneal urological surgeries.

Le texte complet de cet article est disponible en PDF.

Plan


 Conflict of Interest and Financial Disclosures:This study was conducted with financial and logistical support from Intuitive Surgical, Inc (Sunnyvale, CA). Jihad H. Kaouk certifies that all conflicts of interest, including specific financial interests and relationships and affiliations relevant to the subject matter or materials discussed in the manuscript (eg, employment and/or affiliation, grants or funding, consultancies, honoraria, stock ownership or options, expert testimony, royalties, or patents filed, received, or pending), are the following: Endocare, Inc, Intuitive. – J.H. Kaouk (consultant).
Juan Garisto and Riccardo Bertolo have no industrial relationship to disclose.
Funding/Support and Role of the Sponsor:This research project was supported by Intuitive Surgical (Sunnyvale, CA) who provided financial and material support.


© 2018  Elsevier Inc. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 119

P. 91-96 - septembre 2018 Retour au numéro
Article précédent Article précédent
  • Robotic-assisted Urethra-sparing Simple Prostatectomy Via an Extraperitoneal Approach
  • Ping Wang, Dan Xia, SunYi Ye, DeBo Kong, Jie Qin, TaiLe Jing, YeQing Mao, HongZhou Meng, Shuo Wang
| Article suivant Article suivant
  • Low-intensity Extracorporeal Shock Wave Therapy for Erectile Dysfunction: A Systematic Review and Meta-analysis
  • Libo Man, Guizhong Li

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Elsevier s'engage à rendre ses eBooks accessibles et à se conformer aux lois applicables. Compte tenu de notre vaste bibliothèque de titres, il existe des cas où rendre un livre électronique entièrement accessible présente des défis uniques et l'inclusion de fonctionnalités complètes pourrait transformer sa nature au point de ne plus servir son objectif principal ou d'entraîner un fardeau disproportionné pour l'éditeur. Par conséquent, l'accessibilité de cet eBook peut être limitée. Voir plus

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2026 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.