Electron Microscopy and Elemental Analysis of Defective Drysdale Nucleus Manipulator Instruments Implicated in Surgical Complications - 13/12/22

Doi : 10.1016/j.ajo.2022.09.001 
Lloyd R. Kopecny 1, , Joanna M. Biazik 2, Minas T. Coroneo 3
1 From the School of Clinical Medicine, University of New South Wales Sydney (L.R.K.) 
2 Electron Microscope Unit, Mark Wainwright Analytical Centre, University of New South Wales Sydney (J.M.B.) 
3 and Department of Ophthalmology, Prince of Wales Hospital (M.T.C.), Sydney, NSW, Australia 

Inquiries to Lloyd R. Kopecny, Randwick, New South Wales, AustraliaRandwick, New South WalesAustralia

Highlights

Reusable Drysdale manipulators are widely used in cataract surgery.
Damage or residues can accumulate with instrument reuse.
Scanning electron microscopy found sharp defects in all (5 of 5) reused instruments.
Defects had elemental composition mainly of carbon or aluminum.
Sharp defects may contribute to complications including posterior capsule rupture.

Le texte complet de cet article est disponible en PDF.

Résumé

PURPOSE

To analyze microtopography of 5 reusable Drysdale nucleus manipulator (DNM) paddled tips for sharp defects and evaluate their elemental composition to determine probable source, investigating 2 instruments (DNM 1 and 4) implicated in causing posterior capsule rupture (PCR) and 3 instruments with sharp edges identified by finger-tip interrogation intraoperatively.

DESIGN

Experimental laboratory investigation.

METHODS

DNM paddled tips were analyzed using scanning electron microscopy (SEM) to evaluate for sharp surface defects (number, dimensions), and subsequently energy dispersive x-ray spectroscopy (EDS) performed on sharp defects to determine their elemental composition.

RESULTS

All reused DNMs analyzed (5 of 5) had significant structural defects on SEM analysis including sharp burrs, cavities and indentations, surface debris or residues, and roughening, compared to the new instrument (DNM 3, control) which had no defects. DNM 1 had 2 sharp defects, a larger 14 × 76-µm one and a craterlike 167 × 220-µm defect containing debris. EDS found that DNM 2 had 3 of 4 burrs composed mainly of carbon, the fourth of mixed composition (calcium, sulfur, oxygen); DNM 4 had 1 small burr, EDS significant for carbon; DNM 5 had 3 prominent burrs, the largest of 20 × 28 µm, 2 composed of aluminum, and some carbon residue. DNM 6 had 1 burr composed of aluminum and 3 prominent cavity defects, the largest covering 781 µm2.

CONCLUSION

Reusable DNMs are widely used in cataract surgery. Sharp carbon- or aluminum-containing burrs were detected on all reused instruments analyzed together with 1 burr of mixed composition, originating from (1) organic residues, (2) instrument fragments, or (3) salt and contaminant deposits. Sharp defects may contribute to capsular damage including PCR, and residues may pose other safety concerns. Therefore, we support development of a quality, reliable single-use alternative instrument and further encourage careful inspection of all reusable instruments principally by finger-tip interrogation for sharp edges preuse.

Le texte complet de cet article est disponible en PDF.

Plan


 Supplemental Material available at AJO.com.


© 2022  Elsevier Inc. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 245

P. 102-114 - janvier 2023 Retour au numéro
Article précédent Article précédent
  • Full-Thickness Macular Hole: Are Supra-RPE Granular Deposits Remnants of Photoreceptors Outer Segments? Clinical Implications
  • Andrea Govetto, Daniela Bacherini, Mario R. Romano, Martina Matteucci, Sara Lucchini, Lorenzo Cifarelli, Francesco Dragotto, Matteo Cereda, Salvatore Parrulli, Giovanni Staurenghi, Fabrizio Giansanti, Gianni Virgili, Ferdinando Bottoni, Paolo Radice
| Article suivant Article suivant
  • Large Internal Limiting Membrane Tears: Prevalence, Clinical Characteristics, and Surgical Utility
  • Asad F. Durrani, Robert A. Hyde, Mark W. Johnson

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Elsevier s'engage à rendre ses eBooks accessibles et à se conformer aux lois applicables. Compte tenu de notre vaste bibliothèque de titres, il existe des cas où rendre un livre électronique entièrement accessible présente des défis uniques et l'inclusion de fonctionnalités complètes pourrait transformer sa nature au point de ne plus servir son objectif principal ou d'entraîner un fardeau disproportionné pour l'éditeur. Par conséquent, l'accessibilité de cet eBook peut être limitée. Voir plus

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2026 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.